Test Port Output Power Linearity - Agilent Technologies 8753ET Service Manual

Network analyzer
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests

4. Test Port Output Power Linearity

This test verifies the analyzer's test port output power linearity throughout the analyzer's
frequency range. A power meter calibration is used to set the reference output power
across the frequency range of the analyzer. The analyzer's B-channel receiver is used to
determine power linearity after each change in output power.
This test does not verify the functionality of the step attenuator. To verify its
NOTE
function, go to
Analyzer warm-up time: 30 minutes
Specifications
Power Range
−15 to +5 dBm
+5 to +10 dBm
b
+5 to +8 dBm
a. Relative to 0 dBm from 300 kHz to 3 GHz
(2 GHz for Option 075; 6 GHz for Option 006).
b. Options 014 and 075
Required Equipment
Description
Power Meter
Power Sensor
Adapter: APC-7 to type-N (f)
Attenuator: APC-7, 50Ω, 20 dB
RF Cable: APC-7, 50Ω
GPIB Cable
Additional Equipment for Option 006 Analyzers
Power Sensor
Additional Equipment for Option 075 Analyzers
Attenuator: 20dB, Type-N
Minimum Loss Pads (2): 50Ω to 75Ω
Cable: 50Ω, Type-N(m) to Type-N (m)
Power Sensor
2-22
"Step 2. Operator's Check" on page
a
Power Level Linearity
±0.2 dB
±0.5 dB
±0.5 dB
HP/Agilent Part or Model Number
437A/438A or E4418B/4419B
8482A
11524A
8492A Option 020
8120-4779
10833A/B/C/D
8481A
8491A Option 020
11852B
8120-4781
8483A (Option H03)
4-6.
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

8753es

Table of Contents