Test Port Output Power Range And Linearity - Agilent Technologies 8702D Installation Manual

Lightwave component analyzer
Table of Contents

Advertisement

4 Press MENU, CW FREQ, 300, k/m. Set the calibration factor on the power meter for
5 Write the power meter reading on the "Performance Test Record."
6 Repeat
In Case of
Difficulty
1 Be sure the source power is switched on. Press MENU, POWER. Check the SOURCE
2 Refer to the Agilent 8753D Service Guide for more troubleshooting
Required
Equipment
Procedure
1 Zero and calibrate the power meter. Refer to the power meter operating and
Performance Tests

4. Test Port Output Power Range and Linearity

NOTE
The factory preset test port power is 0 dBm.
this CW frequency.
Step 4
and
Step 5
Test Record."
PWR softkey, ON should be highlighted. Otherwise, press SOURCE PWR to switch
on the source power.
information.
4. Test Port Output Power Range and Linearity
Perform this test to verify the analyzer's test port output power range and
power level linearity at selected CW frequencies.
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11524A
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8481A
Additional Required Equipment for Analyzers with Option 075
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8483A Option H03
4-8
for each CW frequency listed in the "Performance