Gpib Interface; Usb Interface; Lan Interface - Agilent Technologies Switch/Measure User Manual

Multifunction switch/measure unit
Table of Contents

Advertisement

GPIB Interface

Each device on the GPIB (IEEE- 488) interface must have a unique address.
You can set the instrument's address to any value between 0 and 30.
The address is set to "9" when the instrument is shipped from the factory.
• Your computer's GPIB interface card has its own address. Be sure to
• The GPIB address is stored in non- volatile memory, and does not
Front Panel Operation: Utility > REMOTE I/O > GPIB > GPIB ADDRESS
To set the GPIB address, turn the knob (or use the number keypad)
to select the desired address.
Remote Interface Operation:

USB Interface

For the USB interface, no configuration parameters are required to set up
the instrument. Connect your instrument to a USB port on your computer.
Note that it may take several seconds for the computer to recognize and
establish a connection to the instrument.

LAN Interface

By default, DHCP is enabled on the instrument, which may enable network
communication over the LAN interface. You may need to set several
configuration parameters as described in this section.
34980A User's Guide
avoid using the computer's address for any instrument on the interface
bus.
change when power has been off, after a Factory Reset (
command), or after an Instrument Preset (
SYSTem:PRESet
SYSTem:COMMunicate:GPIB:ADDRess
Features and Functions
*RST
command).