The Actiwave Cardio has additional settings related to the
in-built 3-axis accelerometer:
The accelerometer sample rate options will vary depending
upon the ECG sample rate. Please consult section 11
'FAQ and Troubleshooting' for details of recording times.
The EMG setup offers the option of setting a low or high
range as EMG signals can vary considerably depending
upon electrode type and positioning.
Refer to section 8 for additional information regarding EMG
recordings.
Click on the 'OK' button to store the setup to the Actiwave
unit. The software will confirm when it is configured and
show the programmed start time.
The Actiwave can now be removed and attached to the
subject. If a long start delay has been set the Actiwave will
enter a low power sleep mode until it needs to start
logging.
Before set-up, the Actiwave memory must be erased - this
can take a few minutes to complete depending on the
Required – Enable the axes of
orientation and select the accelerometer
sample rate.
Required – Select EMG Range:
'Low' (500uV p-p)
'High' (8mV p-p)
Page 19