View System Information; Sorting Setup; Parameter A/B Setup; Nominal Value Setup - UNI-T UTR2810E Series User Manual

Lcr meter
Table of Contents

Advertisement

3.3.10 View System Information

After view the system information, press【ENTER】or【ESC】to exit the system information interface.

3.4 Sorting Setup

Deviation mode is used to control the display way of the result, deviation display is to show the difference between measured
value and reference value. It is used for observing test result with different test conditions or environmental conditions. And it
can also be used to observing the test result when sorting the measurement (set the deviation display mode and deviation
reference to consistent with the deviation tolerances mode and the nominal value).
Press【Shift】and then press【UTIL】to enter sorting setting, the setting method is the same with measurement parameter
setting.

3.4.1 Parameter A/B Setup

Set parameter A/B of UTR2810E/2811E in the main interface, it will keep the settings.

3.4.2 Nominal Value Setup

The nominal value of primary parameter of UTR2810E/2811E can separate set and save.

3.4.3 Comparison Function

The inner comparison of UTR2810E/2811E are allowed to set 3 sets of the limit of primary parameter and 1 set of the limit
of secondary parameter. The measured element can be divided into 5 levels (NG, BIN1, BIN2, BIN3 and AUX). When the
primary parameter use the comparison function, it should turn on 【Comparsion】switch, this function is only charge for
the sorting judgement of primary parameter; When the secondary parameter use comparison function, it should turn on
【Auxiliary】switch, this function is only charge for the sorting judgement of secondary parameter.
Save Setting
Description
ON
Turn on auto save
OFF
Turn off auto save
Figure 3-3 Sorting Interface
Figure 3-4 Sorting Result
20
UTR2810E Series LCR Meter – User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Utr2810eUtr2811e

Table of Contents