6.5 Test Terminal Mode

Four-terminal test:
Hcur:current excitation high end;
Hpot:voltage sampling high end;
Lpot:voltage sampling low end;
Lcur:current excitation low end.

6.6 Test Speed

Test frequency, integral time, component size, display mode, rang mode and comparsion all these will affect test
spee.
UTR2810E/2811E provides thress test modes FAST,MED and SLOW . In general, the slower the test speed, the
more stable and accurate the test results .

6.7 Basic Accuracy

Type
C
L
Z
R
D
Q
Notices:
1、D,Q is absoult error, other is relative erro Dx=1/Qx;
2、The subscript x is the measured value of this parameter, the subscript max is the
maximum value and min is the minimum value;
3、 ks is speed factor, kv is voltage factor, kf if frequency factor, ke is temperature factor(see
Table 6-7);
4、In order to ensure the measurement accuracy, accuracy calibration should under the
current measurement condition, test tool to perform the stable open-circuit zero
clearing;
5、Guarantee of accuracy:1 year
Speed
Time(S)
FAST
23.3
5.4
MED
2.8
SLOW
Table 6-3 Test Speed
Accuracy
0.1% (1+ Cx/Cmax+ Cmin/Cx )(1+Dx )( 1+ks+kv+kf )ke
0.1% (1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx )( 1+ks+kv+kf )ke
0.1% (1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf )ke
0.1%(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf )ke
0.0010(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx2 )(1+ks+kv+kf )ke
0.0015(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf )ke
Table 6-4 Basic Accuracy
31
UTR2810E Series LCR Meter – User Manual