1.2 Configuring the KDR panel

1. In Ultra, select the Device Configuration tab.
2. Under the Configuration Pool section, select
Default, then select KDR (Figure 3).
3. Under the device settings, use the following
configuration values:
Field
Maven
Exposure Index
Edge Crop
AEC Display
Global Orientation
Image Processing
X-Ray
Synchronization
XG-IF Box ID
Exposure Time
Figure 3 Device configuration
VZ0099HTG201606-1.1
Value
Aero, 32 Bit, #1
Determined during
installation
(Default is 1000)
Default value, 0
Default value, 0
Default value, 0
Default value, 0
Integrated
1
1.0
Customer Support
Ultra KDR Panel Configuration and Calibration Guide v1.1

1.3 Panel Registration

1.3.1 Before you begin

You will need to power on the KDR panel and have
the device identity list available (Figure 4), which is
provided in the IF cable box. The device identity list
includes information such as the serial number and
Sub device of the IF cable.
For detailed device information refer to KDR System Operation Manual
(A9KXBB01EN01_160316) Chapter 2.
Figure 4 Device identity list
Also on the front of the power box (XG-IFBox.), flip
the switch to the on position as shown in (Figure 5)
Figure 5 XG-IFBox
After switch is flipped into the on position, please allow
up to 30 seconds for the panel to start up and initialize.