KDR Panel Registration, Configuration and
Calibration
This section describes the configuration of the KDR panel in Ultra software.

1.1 Before you begin

For Panel Specification, Safety Information and Handling instructions
please see KDR System Operation Manual
(A9KXBB01EN01_160316).
Before starting panel setup, verify the network
settings for the onboard card.
The following configuration applies to Ultra 4.1.0.x
or later with the KDR Panel and driver 1006. If you
have with anything prior to Ultra 4.1.0, contact your
sales vendor.
1. Open Network Connections.
2. Right-click on the onboard network card and
select Properties.
3. Click to highlight Internet Protocol Version 4
(TPC/IPv4) and select Properties.
Figure 1 Onboard network card
VZ0099HTG201606-1.1
Ultra KDR Panel Configuration and Calibration Guide v1.1
Customer Support
4. Set the following addresses:
IPV4 address:
Subnet mask:
Figure 2 IPv4 address configuration
1.800.366.5343 –
HCITsupport@konicaminolta.com
Ultra
192.168.20.100
255.255.255.0
3