Agilent Technologies E1439 User Manual page 190

Vxi 70 mhz if adc with filters and memory
Table of Contents

Advertisement

Agilent E1439 Programmer's Reference
Functions listed alphabetically
age1439_sync_output
Selects the output for the sync signal. This description also includes the query function:
VXIplug&play Syntax
ViStatus age1439_sync_output(ViSession id, ViInt16 syncOutput);
ViStatus age1439_sync_output_get(ViSession id, ViPInt16 syncOutputPtr);
Note
This command should be used only for specialized custom clock requirements. Most useful clock
setups can be supplied by age1439_clock_setup.
Description
This function selects which output the module should use for it's sync signal.
Parameters
id
is the VXI instrument session pointer returned by the age1439_init function.
syncOutput
AGE1439_SYNC_OUT_OFF
AGE1439_SYNC_OUT_BOTH
SMB sync connectors and the VXI backplane.
AGE1439_SYNC_OUT_SMB
sync connectors.
AGE1439_SYNC_OUT_VXI
syncOutputPtr
Returns a pointer to the current value of syncOutput.
Return Value
AGE1439_SUCCESS indicates that a function was successful.
Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to
page
See Also
"Default values" on page
"age1439_sync_clock" on page
sync" in chapter 3
180
age1439_sync_output_get
#include "age1439".h
102.
201,
specifies no sync signal output.
specifies that the sync signal be output to both the front panel
specifies that the sync signal be output to the front panel SMB
specifies that the sync signal be output to the VXI backplane.
"age1439_init" on page
132,
178,
"age1439_sync_direction" on page
"age1439_error_message" on
"age1439_clock_setup" on page
179,
"Using clock and
78,

Advertisement