age1439_front_panel_clock_input
Specifies the source for the front panel clock. This description also includes the query function:
VXIplug&play Syntax
ViStatus age1439_front_panel_clock_input(ViSession id, ViInt16 fpClock);
ViStatus age1439_front_panel_clock_input_get(ViSession id, ViPInt16 fpClockPtr);
Note
This command should be used only for specialized custom clock requirements. Most useful clock
setups can be supplied by age1439_clock_setup.
Description
This function selects a front panel clock source that is used to drive the analog to digital converter
(ADC) for single module operation or when a module is used as the master ADC clock source for
a multi-module system.
Parameters
id
is the VXI instrument session pointer returned by the age1439_init function.
fpClock
AGE1439_CLOCK_OFF
AGE1439_SMB_CLOCK
connectors.
AGE1439_BNC_CLOCK
connector.
fpClockPtr
returns a pointer to the current value of fpClock.
Return Value
AGE1439_SUCCESS indicates that a function was successful.
Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to
page
See Also
"Commands which halt active measurements" on page
"age1439_init" on page
chapter 3
age1439_front_panel_clock_input_get
#include "age1439".h
specifies no front panel source.
specifies clock input from the front panel Intermodule Clock/SMB
specifies clock input from the front panel Ext Clock/Ref BNC
102.
132,
"age1439_clock_setup" on page
Agilent E1439 Programmer's Reference
Functions listed alphabetically
"age1439_error_message" on
198,
"Default values" on page
78,
"Using clock and sync" in
201,
131