58 | DX-D 60C, DX-D 60G | Introduction to the DR Detector

Limitations for patient contact

This device is not intended to supply heat to a patient. However, during
normal use, surfaces will become heated due to power dissipation.
Patient contact surfaces will not exceed 48 degrees C under normal use
conditions. The operator should monitor and evaluate how much of the
patient's body area is in contact with these surfaces and for how long.
If the temperature limit is exceeded, an error is displayed and further
exposures are impossible to avoid further heating, until the temperature has
dropped.
0294B EN 20190221 1131