3 Setup

Setup of this system is normally executed by our service personnel. This section describes setup procedures
briefly.

3-1 Cable connections

NOTE
NOTE
TIP
TIP
1.
Connect LD-LDPSU inline cable to FV5-LD405 and fasten cable pull-stopper firmly.
Fasten screws of cable pull-stopper tightly for sure. In case that LD-LDPSU
inline cable comes off from FV5-LD405, FV5-LD405 becomes shaky and it would
lead to breakage. In addition, if cable comes off during running time, it would
cause a failure.
2.
Connect LD-LDPSU inline cable to LD-HEAD connector of FV5-LDPSU and fasten
cable pull-stopper firmly.
Fasten screws of cable pull-stopper tightly for sure. In case that LD-LDPSU
inline cable comes off from FV5-LDPSU, FV5-LD405 becomes shaky and it would
lead to breakage. In addition, if cable comes off during running time, it would
cause a failure.
3.
Insert power cord into power cord connector of FV5-LDPSU.
When you use FV300 or FV500, connect FV-COMB cable from FV5-COMB(2) or FV5-
COMBA with FV5-LDPSU. Operation of FLUOVIEW software from FV300 or FV500 is
possible. For detail, refer to FV300 or FV500 manual.
When you use FV300 or FV500 and you set AOTF combiner (FV5-COMBA), you can
use REX function at the following connections and settings.
1.
Connect BLANKING connector of FV5-LCU and ANALOG connector of
FV5-LDPSU with BNC cable.
2.
Connect UV connector of FV5-LCU and DIGITAL connector of FV5-LDPSU with
BNC cable.
3.
Set UV rotary switch of FV5-LCU to "1".
For details of REX function see instruction manual of FV300 or FV500.
Setup
/ Cable connections