Testing Dc Current - Fluke 525A Service Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525A:
Table of Contents

Advertisement

Testing DC Current

Use the Wavetek 1281 and the precision shunt to measure the 525A output as shown in
Figure 7. Take the Voltage reading from the Wavetek 1281 and divide it by the 742A-1
actual value. See Table 12.
525A
525A
TEMPERATURE / PRESSURE CALIBRATOR
mA
VOLTS
RTD
100mA MAX
HI
HI
OUTPUT
100V MAX
TC
LO
LO
INPUT/OUTPUT
20V PK
MAX
4W RTD
HI
20V PK
MAX
INPUT
SENSE
LO
20V PK
MAX
mA
Output
Terminals
100 mA
Output (A)
Volt
0.000
0.025
0.075
0.100
Current Shunt
STBY
VOLTS
TC
TYPE
OPR
mA
RTD
UNITS
OUTPUT
INPUT
ZERO
7
8
9
SHIFT
SETUP
CJC
C / F
4
5
6
SET
RECALL
AUTOSET
1
2
3
RNG LOCK
LOCAL
EXP
/
0
ENTER
CE
Figure 7. Measuring DC Current
Table 12. Measuring DC Current
Current
(I=E/R)
1281 DCV Function
742A-1
Shunt
Value
90 Day (A)
Direct Into
0.000002
1281 Input
0.0000041
0.0000083
0.0000105
525A
Performance Tests
ajr25f.eps
1 Year (A)
0.000002
0.0000045
0.0000095
0.000012