4 - 16
Checking the tube current
•
Connect oscilloscope on D910 to MP "IRS" (NOMINAL), "IR" (ACTUAL) and "GND".
•
POLYMOBIL ON
•
Trigger an exposure with the default values 60 kV, 10 mAs.
IRS
IR
POLYMOBIL III
Notes on trouble-shooting
RXB8-115.061.02
Rev. 01
10.95
Page 16 of 24
CS PS 24
4
Siemens AG
Medical Solutions