Checking The High-Voltage Kv Nom And Kv Act - Siemens POLYMOBIL III Service Instructions Manual

Hide thumbs Also See for POLYMOBIL III:
Table of Contents

Advertisement

4 - 12
Checking the high-voltage kV
Connect oscilloscope on D910 to MP "kVS" (1 V
10 kV actual value) and "GND" (X038I-6).
POLYMOBIL ON
Trigger an exposure with the default values 60 kV; 10 mAs.
POLYMOBIL III
Notes on trouble-shooting
and kV
NOM
20 kV nominal value) and "kV" (1 V
RXB8-115.061.02
Rev. 01
10.95
ACT
Page 12 of 24
CS PS 24
4
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

Table of Contents