Checking the supply voltages

When the system is switched on, relays HR and K1 on D920 pull in and supply
transformer T1, power supply M6 and the intermediate circuit with line voltage.
Measure the 24 V supply voltage with the digital multimeter on D920 at MP "+ 24 V" and
"0V"..
Measured voltage approx. 28.2 V
Measure the filament circuit voltage supply on the electrolyte capacitor
C46 at +/ - on D920.
Measured voltage
in stand-by mode approx. 18.5 V
with preparation approx. 16.4 V
POLYMOBIL III
Notes on trouble-shooting
RXB8-115.061.02
Rev. 01
10.95
D920 (X038I-92D)
±
Tolerance
5 V
±
Tolerance
4 V
Tolerance
±
3 V
Page 6 of 24