Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
NOTE NOTE is used to indicate information which explains the proper way to use devices or to carry out a process, i.e. which provides hints and tips. Polymobil 10 RXR8-120.101.01 Page 2 of 14 Siemens AG Rev. 05 07.05...
- Release radiation for as short a time as possible. Checks in which radiation must be released are identified by the radiation warning symbol. Siemens AG RXR8-120.101.01 Page 3 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 8
1 - 4 General information Fig. 1 Polymobil 10 RXR8-120.101.01 Page 4 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
Page 9
Only connect the POLYMOBIL 10 to a power supply (wall outlet) whose installation corresponds to the requirements according to VDE 0107 or the national regulations. • Prior to any intervention, switch the POLYMOBIL 10 off using the power OFF switch on the control console, and disconnect the power plug. •...
During the measurement, move the power cable and additional connection cables with an integrated protective conductor section by section to detect cable breaks. The protective conductor resistance must not exceed 0.2 Ohms. Polymobil 10 RXR8-120.101.01 Page 6 of 14 Siemens AG Rev.
Page 11
"Certificates". If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Siemens AG RXR8-120.101.01 Page 7 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
However, please note the minimum resolution of the leakage current measuring instru- ment and any additional manufacturer's data restricting the use of the measuring device. Polymobil 10 RXR8-120.101.01 Page 8 of 14 Siemens AG Rev.
Page 13
DIN VDE 0751-1/2001-10, Fig. C5 for protection class 1 = System 2 = Applied part, type B (if present) 3 = Measuring arrangement (integrated in measuring instrument) Siemens AG RXR8-120.101.01 Page 9 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 14
Separate this report, fill it out, and file it in the "Certificates" register in the system manual. If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Polymobil 10 RXR8-120.101.01 Page 10 of 14 Siemens AG Rev.
2.9 Radiation; tests: the devices present and not damaged? SIE Radiation indicator (no radiation protection of the building) SIE Acoustic signal Tab. 1 Siemens AG RXR8-120.101.01 Page 11 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 16
SIM Support arm locking mechanism (lubricated, if required)? 2.6 Single tank; check: SIM Locking mechanism 2.7 Double-slot diaphragm, check: SIM Rotatability Tab. 1 Polymobil 10 RXR8-120.101.01 Page 12 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
Page 17
(e.g., safety floor plate)? 5.7 Are the positions of the table and n.a. system reproducible (e.g., zero positions, layer height, etc.)? Tab. 1 Siemens AG RXR8-120.101.01 Page 13 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 18
Have the operator, or a person authorized by the operator, sign the TSC report. Hand the TSC report over to the operator, or a person authorized by the operator. Tab. 1 Polymobil 10 RXR8-120.101.01 Page 14 of 14 Siemens AG Rev. 05 07.05...
Make sure that the handles are secure and tighten them if necessary: Single tank holder Single tank Control handle over the control console Siemens AG RXR8-120.101.01 Page 1 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 20
2 - 2 Inspection and maintenance NOTE If screws or securing elements are replaced, the screw locks must also be replaced. Polymobil 10 RXR8-120.101.01 Page 2 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
• Pedal positions as shown in 3/Fig. 1 The unit can be moved forward or backward. It is steered through the front wheels. Siemens AG RXR8-120.101.01 Page 3 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
- Remove the clamp handle screw (4/Fig.3), and use the screwdriver to press the clamp handle away from the axle. - Attach the clamp handle in the vertical position and tighten the screw (4/Fig.3). - Recheck the locking mechanism of the single tank. Polymobil 10 RXR8-120.101.01 Page 4 of 14 Siemens AG Rev.
Move the support arm to the top position. • Use the spirit level on the double-slot diaphragm to check the 0° position. If necessary, readjust the scale on both sides. Siemens AG RXR8-120.101.01 Page 5 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Check the edges of the locking disks (6/Fig.6). They must not be worn. • Ensure that there is play (without any friction) on all sides of the locking bar in the stand head (7/Fig.6). Polymobil 10 RXR8-120.101.01 Page 6 of 14 Siemens AG Rev.
Page 25
Remove cover (3/Fig.7) from the stand. • Carefully check the condition of the spring counterweight chains (5/Fig.8). Replace the stand column if the chains are defective. Siemens AG RXR8-120.101.01 Page 7 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Trigger exposure. The radiation indicator (1/Fig.9) must be lit during exposure, and an acoustic signal will sound at the same time. • POLYMOBIL OFF and disconnect the power plug. Polymobil 10 RXR8-120.101.01 Page 8 of 14 Siemens AG Rev. 05 07.05...
Page 27
If the oscillogram ist not produced, perform an adjustment according to the service instructions. • POLYMOBIL OFF and disconnect the power plug. • Close the control console. Siemens AG RXR8-120.101.01 Page 9 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 28
• Make a note on the processed film of the following data using a waterproof felt tip pen: - Adjusted SID - Film size - Radiation field size Polymobil 10 RXR8-120.101.01 Page 10 of 14 Siemens AG Rev. 05 07.05...
Check the operating elements and indicators on the control console. - See "Overview of the Operating Elements" in the operator manual. • POLYMOBIL OFF and disconnect the power plug. Siemens AG RXR8-120.101.01 Page 11 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Separate this report, fill it out, and file it in the system manual under "Certificates". If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Polymobil 10 RXR8-120.101.01 Page 12 of 14 Siemens AG Rev.
"Certificates" register in the system manual. If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Siemens AG RXR8-120.101.01 Page 13 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Do not use scouring cleaning agents or organic solvents or solvent-based cleaning agents (for example, benzine, alcohol, spot remover) because of potential incompatibility with the material! Do not spray the POLYMOBIL! Never let cleaning agents get into the POLYMOBIL! Polymobil 10 RXR8-120.101.01 Page 14 of 14 Siemens AG Rev.
Meas. inst. Ser. No.: Meas. inst. calibrated up to: Assessment: Date: Name: Signature: (*1) Meas. circuit: See Fig. 2 to Fig. 3, next page Siemens AG RXR8-120.101.01 Page 3 of 6 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 36
DIN VDE 0751-1:2001-10, Fig. C6 for protective class 1 = System 2 = Application part (if present) 3 = Measuring arrangement (integrated in the measuring instrument) Polymobil 10 RXR8-120.101.01 Page 4 of 6 Siemens AG Rev. 05 07.05...
Page 37
Electrical safety/reports 3 - 5 Comments: Date Remarks Name Signature Siemens AG RXR8-120.101.01 Page 5 of 6 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 38
3 - 6 Electrical safety/reports Date Remarks Name Signature Polymobil 10 RXR8-120.101.01 Page 6 of 6 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
Page 39
Protective conductor test Completely rewritten maintenance Inspection and Leakage current Completely rewritten maintenance measurement Inspection and Concluding work "Disinfection" checkpoint deleted maintenance Electrical n.a. Chapter added safety/reports Siemens AG RXR8-120.101.01 Page 1 of 2 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 40
4 - 2 Changes to Previous Version Polymobil 10 RXR8-120.101.01 Page 2 of 2 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
Page 42
The Maintenance Protocol is a document and must therefore be archived. After comple- ting maintenance, it must be filed in the appropriate register of the Maintenance binder. If needed, give the customer a copy of it. Polymobil 10 RXR8-120.105.01 Page 2 of 8 Siemens AG Rev.
Page 43
For safety reasons, continued operation of the system is possible only after the deficiencies have been successfully corrected. Town: Date: Technician's name: Signature: Siemens AG RXR8-120.105.01 Page 3 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 44
CS No.: Date: Activities performed If there is no entry in the table, see the entries beginning on Page 7. Date: Activities performed: not OK n.a. Polymobil 10 RXR8-120.105.01 Page 4 of 8 Siemens AG Rev. 05 07.05 CS PS 24...
Page 45
Preventive Maintenance, Preventive Parts Replacement, Visual Inspection Preventive maintenance adjustments Preventive Check of Operating Values/Functions Quality Check Image Quality Check System Quality Software Maintenance Customer Service Engineer Siemens AG RXR8-120.105.01 Page 5 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 46
If 'Yes', enter the component (including Serial No.) into the table that was used for mainte- nance. Movable components (including test phantoms that are part of the system) are parts that can be used on different systems. Movable Component Serial No. Polymobil 10 RXR8-120.105.01 Page 6 of 8 Siemens AG Rev. 05 07.05...
Page 47
2 - 5 Light localizer lamp 2 - 5 Rotatability 2 - 5 Format setting 2 - 5 0° position 2 - 5 Siemens AG RXR8-120.105.01 Page 7 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
Page 48
3 - 1 3.1. Protective conductor resistance/report 3 - 1 3.2. System leakage current/report 3 - 3 Changes to previous version 4 - 1 Polymobil 10 RXR8-120.105.01 Page 8 of 8 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...