The obtained maximum density (Dmax (3.20)) after calibration is
displayed:
• Calibration successful, but maximum density (Dmax) differs more
than 2% from the target value.
The obtained maximum density (3.02) after calibration is displayed:
• Calibration successful, but the maximum density has been reached.
3.02
The obtained maximum density (2.95) after calibration is displayed:
5. Press the Confirm key to return to the Calibration menu.
2.95
6. Check the printed test film.
The test film should be similar to the image below (e.g. 14x17" film
format).
Related Links
Accessing the Calibration menu
Drystar AXYS | Advanced Operation (key-operator mode) | 65
3.02
3.02
2.95
2.95
WARNING:
When the calibration process is interrupted, you will have to
start the calibration procedure again. Otherwise, the old
calibration parameters will be used.
Film Calib53_14x 17.BMP ( Ver sion 2)
Film Calib53_14x 17.BMP ( Ver sion 2)
Film Calib53_14x 17.BMP ( Ver sion 2)
WARNING:
The calibration film must be free of dust particles or any
other artifacts. If this is not the case, clean the print head
and restart the calibration procedure.
on page 63
2853E EN 20210520 0952