Agilent Technologies PHD-4 User Manual page 84

Portable helium detector
Hide thumbs Also See for PHD-4:
Table of Contents

Advertisement

5
Technical Information
Troubleshooting
Trouble
Unstable reading/Reading Drift
Trouble
Low sampled slow
132/140
Answer
Due to saturated sensor (spikes) or high room
temperature (drift)
Operate with environmental temperature
between 5 °C and 35 °C - 90 %HR
Answer
Due to clogging of line
Check probe line (pipe)
Replace glass-wool filter
Clean sintered filter
Start Sampled flow auto-adjust in addition to periodical
start (20h)
PHD-4 Portable Helium Detector User Manual / 87-900-120-01 (F )