Brief Operating Instructions
Flowmeter
Proline 10
Modbus RS485 transmitter
Transmitter with Coriolis sensor
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline 10

  • Page 1 Services KA01500D/06/EN/01.21 71510510 2021-07-01 Brief Operating Instructions Flowmeter Proline 10 Modbus RS485 transmitter Transmitter with Coriolis sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 2 of 2: Transmitter Contain information about the transmitter.
  • Page 2 Flowmeter Proline 10 Modbus RS485 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 Flowmeter Proline 10 Modbus RS485 Brief Operating Instructions for flowmeter Brief Operating Instructions for flowmeter The device consists of a transmitter and a sensor. The process of commissioning these two components is described in two separate manuals that together form the Brief Operating Instructions for the flowmeter: •...
  • Page 4: Table Of Contents

    Table of contents Flowmeter Proline 10 Modbus RS485 Table of contents About this document ............. . 5 Symbols .
  • Page 5: About This Document

    Flowmeter Proline 10 Modbus RS485 About this document About this document Symbols 1.1.1 Warnings DANGER This symbol alerts you to an immediate dangerous situation. Failure to avoid the situation will result in a fatal or serious injury. WARNING This symbol alerts you to a potentially dangerous situation. Failure to avoid the situation may result in a fatal or serious injury.
  • Page 6: Safety Instructions

    Safety instructions Flowmeter Proline 10 Modbus RS485 Forbidden procedures, processes or actions Additional information Reference to documentation Reference to page Reference to graphic Measure or individual action to be observed Series of steps 1. 2. ,… Result of a step...
  • Page 7: Adhesive Labels, Tags And Engravings

    Flowmeter Proline 10 Modbus RS485 Safety instructions Adhesive labels, tags and engravings ‣ Pay attention to all the safety instructions and symbols on the device. Environment and process ‣ Only use the device for the measurement of appropriate media. ‣...
  • Page 8: Product Information

    Product information Flowmeter Proline 10 Modbus RS485 ‣ Only use original spare parts and original accessories from Endress+Hauser. Product information Designated use The device is intended only for the flow measurement of liquids and gases. Depending on the version ordered, the device measures potentially explosive, flammable, poisonous and oxidizing media.
  • Page 9: Product Design

    Flowmeter Proline 10 Modbus RS485 Installation Product design The transmitter and sensor form a mechanical unit. A0043525  1 Main device components Housing cover Display module Transmitter housing Sensor Installation For detailed information about installing the sensor, see the Brief Operating Instructions for the sensor →...
  • Page 10: Turning The Transmitter Housing

    Installation Flowmeter Proline 10 Modbus RS485 Turning the transmitter housing Loosen the fixing screws on both sides of the transmitter housing. NOTICE Overrotation of the transmitter housing! Interior cables are damaged. ‣ Turn the transmitter housing a maximum of 180° in each direction.
  • Page 11: Post-Installation Check

    Flowmeter Proline 10 Modbus RS485 Installation Post-installation check Is the device undamaged (visual inspection)?  Does the device comply with the measuring point specifications? For example: • Process temperature  • Process pressure • Ambient temperature • Measuring range Has the correct orientation been selected for the device? ...
  • Page 12: Electrical Connection

    Electrical connection Flowmeter Proline 10 Modbus RS485 Electrical connection Connection conditions 5.1.1 Notes on the electrical connection WARNING Components carry voltage! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Have electrical connection work carried out by appropriately trained specialists only.
  • Page 13 Flowmeter Proline 10 Modbus RS485 Electrical connection 5.2.2 Permitted temperature range • Observe the installation guidelines that apply in the country of installation. • The cables must be suitable for the minimum temperatures and maximum temperatures to be expected. 5.2.3 Power supply cable (incl.
  • Page 14: Connecting The Transmitter

    Electrical connection Flowmeter Proline 10 Modbus RS485 Connecting the transmitter 5.3.1 Transmitter terminal connections A0045438 Cable entry for power supply cable: supply voltage Cable entry for signal cable Ground terminal, outer 5.3.2 Terminal assignment The terminal assignment is documented on an adhesive label.
  • Page 15 Flowmeter Proline 10 Modbus RS485 Electrical connection NOTICE If the cable gland is incorrect, this compromises the sealing of the housing! Damage to the device. ‣ Use a suitable cable gland corresponding to the degree of protection. Ground the device carefully and provide potential equalization.
  • Page 16 Electrical connection Flowmeter Proline 10 Modbus RS485 The cable must be in the tab for strain relief. Let the display module hang down. A0041354 Remove dummy plug if present. NOTICE If the sealing ring is missing, the housing is not sealed tight! Damage to the device.
  • Page 17: Removing A Cable

    Flowmeter Proline 10 Modbus RS485 Electrical connection Removing a cable Use a flat-blade screwdriver to press down on the slot between the two terminal holes and hold. 3 (0.12) Remove the cable end from the terminal. A0029598  2 Engineering unit mm (in) Ensuring potential equalization No special measures for potential equalization are required.
  • Page 18: Hardware Settings

    Electrical connection Flowmeter Proline 10 Modbus RS485 Hardware settings 5.6.1 Enabling write protection Loosen the Allen key of the securing clamp. Unscrew the housing cover in the counterclockwise direction. 3 mm 1,5 Nm (1.1 lbf ft) A0041094 Press the tab of the display module holder.
  • Page 19: Post-Connection Check

    Flowmeter Proline 10 Modbus RS485 Electrical connection Post-connection check Is the protective earthing established correctly?  Are the device and cable undamaged (visual check)?  Do the cables meet the requirements?  Is the terminal assignment correct?  Are all the cable glands installed, firmly tightened and leak-tight? ...
  • Page 20: Operation

    Operation Flowmeter Proline 10 Modbus RS485 Operation Overview of the operating options XXXXX ṁ 3.23 kg/min m /min ³ 23.0 °C A0044206 Local operation via touch screen Computer with operating tool, e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM Field Xpert SFX350 or SFX370 via Bluetooth...
  • Page 21 Flowmeter Proline 10 Modbus RS485 Operation 6.2.2 Navigation • Open menus. • Select items in a list. • Acknowledge buttons. • Enter characters. Swipe horizontally Display next or previous page. Swipe vertically Display additional points in a list. 6.2.3 Operational display During routine operation, the local display shows the operational display screen.
  • Page 22 Operation Flowmeter Proline 10 Modbus RS485 Quick access Locking status Bluetooth is active. Device communication is enabled. Status signal: function check Status signal: maintenance required Status signal: out of specification Status signal: failure Status signal: diagnostics active. 6.2.4 Quick access The Quick access menu contains a selection of specific device functions.
  • Page 23 Flowmeter Proline 10 Modbus RS485 Operation Enter access code. Write protection is enabled. Back to operational display. 6.2.5 Main menu The main menu contains all the menus required for the commissioning, configuration and operation of the device. Main menu and navigation...
  • Page 24 Operation Flowmeter Proline 10 Modbus RS485 Submenus and navigation Main menu Main menu Comissioning Guidance Device ident. Device tag Commissioning XXXXX Guidance Diagnostics Application Serial number Firmware version Device name System User manag. Language A0044219 • Open the main menu.
  • Page 25 Flowmeter Proline 10 Modbus RS485 Operation XXXXX F180 Temperature exceeded Help A0043008 Device status Diagnostic behavior with diagnostic code Short text Open the troubleshooting measures. 6.2.7 Editing view Editor and navigation The text editor is used to enter characters. Text editor...
  • Page 26 Operation Flowmeter Proline 10 Modbus RS485 • Enter characters. • Select next character set. Swipe horizontally Display next or previous page. Input field Upper case Lower case Numbers Special characters 6.2.8 Date The device has a real-time clock for all log functions. The time can be configured here.
  • Page 27: Smartblue App

    Flowmeter Proline 10 Modbus RS485 Operation 20xx yyyy A0043043 Increase date by 1. Actual value Decrease date by 1. Confirm settings. Cancel editor. • Make settings. • Confirm settings. • Cancel editor. SmartBlue App The device has a Bluetooth interface and can be operated and configured using the SmartBlue App.
  • Page 28: System Integration

    System integration Flowmeter Proline 10 Modbus RS485 Download Endress+Hauser SmartBlue App: • Google Playstore (Android) • iTunes Apple Shop (iOS devices) Supported functions • Configuration of the device • Access to measured values, device status and diagnostic information System integration For detailed information on system integration, see the Operating Instructions for the device.
  • Page 29: Switching On The Device

    Flowmeter Proline 10 Modbus RS485 Commissioning Switching on the device ‣ Switch on the supply voltage for the device.  The local display switches from the start screen to the operational display. XXXXX Volume flow 1767.2063 Starting up... A0042938 If device startup is not successful, the device displays an error message to this effect .
  • Page 30: Commissioning The Device

    Commissioning Flowmeter Proline 10 Modbus RS485 Commissioning the device 8.3.1 Local operation Detailed information on local operation: Via the "Menu" symbol, open the main menu.  Main menu XXXXX 3.23 ṁ kg/min Guidance Diagnostics Application m /min ³ 3.23 Ʃ...
  • Page 31 Flowmeter Proline 10 Modbus RS485 Commissioning Start the Commissioning wizard.  Main menu Comissioning Guidance Device ident. Device tag Commissioning XXXXX Serial number Firmware version Device name A0043018 Follow the instructions on the local display.  The Commissioning wizard goes through all the device parameters that are necessary to commission the device.
  • Page 32: Protecting Settings From Unauthorized Access

    Diagnostics and troubleshooting Flowmeter Proline 10 Modbus RS485 Protecting settings from unauthorized access 8.4.1 Write protection switch Write access to the entire operating menu can be locked via the write protection switch. The values of the parameters cannot be changed. Write protection is disabled when the device leaves the factory.
  • Page 36 *71510510* 71510510 www.addresses.endress.com...