Source container FQG61,
FQG62
Radiometric level measurement
Technical Information and Operating Instructions
Source container with source insert for manual or pneumatic switch-ON/switch-
OFF
Application
The FQG61 and FQG62 source containers are designed to hold the radioactive source
during radiometric point level measurement, continuous level measurement and
density measurement. The radiation is emitted almost unattenuated in one direction
only, and is damped in all other directions. FQG61 and FQG62 differ from each other
in terms of size and screening effect.
Your benefits
• Lightweight device provides best possible screening thanks to almost spherical
design
• Safe and easy source replacement
• Highest safety classification for the source supplied (DIN 25426/ISO 2919, typical
classification C66646)
• Compact device that is easy to mount
• Various angles of emission for optimum adaptation to the application
• Manual or pneumatic switch-ON/switch-OFF
• Padlock, cylinder lock or locking bolt for fixing the switch position
• Switch state easily identified
• Fire-resistant version +821 °C (+1510 °F) / 30 minutes
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser FQG61

  • Page 1 The radiation is emitted almost unattenuated in one direction only, and is damped in all other directions. FQG61 and FQG62 differ from each other in terms of size and screening effect.
  • Page 2: Table Of Contents

    FQG61/FQG62; order code 020, option A ... . FQG61/FQG62; order code 020, option B ... . . 43 Safety instructions ......8 FQG61/FQG62;...
  • Page 3: Product Identification

    Source container FQG61, FQG62 Product identification Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 4: About This Document

    Source container FQG61, FQG62 About this document Symbols used Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 5 Comment TI00439F/00 • Technical Information for gamma radiation source FSG60/FSG61 • Returning source containers • Type A packaging Source container FQG60, FQG61, FQG62, FQG63, FQG66 Documentation Comment SD00297F/00 Instructions for loading and changing the source. Label Clamping device FHG61 Documentation...
  • Page 6 Source container FQG61, FQG62 Gammapilot M FMG60 Documentation Comment TI00363F/00 Technical Information for Gammapilot M FMG60 BA00236F/00 Operating Instructions for Gammapilot M FMG60 (HART) BA00329F/00 Operating Instructions for Gammapilot M FMG60 (PROFIBUS PA) BA00330F/00 Operating Instructions for Gammapilot M FMG60...
  • Page 7 Source container FQG61, FQG62 Certificate of suitability A0037353 Endress+Hauser...
  • Page 8: Safety Instructions

    Safety instructions Designated use The FQG61 and FQG62 source containers described in this document contain the radioactive source, which is used for radiometric measurement of level, interface and density. They screen the radiation from the surrounding environment and allow it to be emitted almost unattenuated in the direction of measurement only.
  • Page 9: General Instructions On Radiation Protection

    Distance Screening Ensure the best possible screening between the radiation source and yourself and all other persons. Source containers (e.g. FQG60, FQG61/FQG62, FQG63, FQG66) and all high-density materials (lead, iron, concrete etc.) can be used for effective screening purposes. Time The time spent in the exposed area should be kept to a minimum.
  • Page 10: Supplementary Safety Instructions

    The limit values for the control zone are specified in the current Radiation Protection Ordinance applicable for your area. The Endress+Hauser sales organization will be pleased to provide further information on radiation protection and regulations in other countries.
  • Page 11: Function And System Design

    Function Function of the radiation source container In the FQG61/FQG62 source container, the radioactive source is surrounded by a steel casing filled with lead which screens off gamma radiation. The radiation can be emitted, almost unattenuated, in one direction only through a channel (focused narrow beam path). This radiation is used for radiometric measurement.
  • Page 12 580 "Test, Certificate". For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components →...
  • Page 13 Source container FQG61, FQG62 Dose rate diagrams for 90° 90° 120° 60° 120° 60° 150° 150° 30° 30° 0° 0° 180° 180° 60% 100% 10% 15% 210° 330° 210° 330° 240° 300° 240° 300° 270° 270° A0018384 FQG61 FQG62 Option in order code...
  • Page 14: Mechanical Construction

    FQG61 251 (9.88) FQG62 272 (10.7) FQG61 The mounting flange of FQG61 and FQG62 is compatible with: DN 100 PN16 (ø 180 mm (7.09 in)) and ANSI 4" 150 lbs (ø 190 mm (7.48 in)) FQG62 FQG61 279 (11) FQG62 360 (14.2)
  • Page 15 Dimensions Version mm (in) Comment FQG61 479 (18.9) Clearance required to replace the radiation source FQG62 560 (22) FQG61/FQG62; order code 020, option B→  54 ø220 (8.66) A0018386  2 Dimensions: mm (in) Dimensions Version mm (in) Comment FQG61 251 (9.88)
  • Page 16 287 (11.3) FQG62 368 (14.5) FQG61 450 (17.7) Clearance required to replace the radiation source FQG62 570 (22.4) FQG61/FQG62; order code 020, option D→  54 ø220 (8.66) A0018388  4 Dimensions: mm (in) Reference O-ring Dimensions Version mm (in)
  • Page 17 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option K, L, M or N→  54 ø220 (8.66) A0018389  5 Dimensions: mm (in) Reference O-ring Dimensions Version mm (in) Comment FQG61 251 (9.88) FQG62 272 (10.7) FQG61 427(16.8) FQG62 508 (20.0)
  • Page 18: Radiation Emission Channel

    According to feature 240 of the product structure: • 5° • 20° • 40° Width of emission • FQG61: 10 mm (0.39 in) • FQG62: 12 mm (0.47 in) Attenuation of the Approx. 0.3 half-value layers (F = 1.2) useful beam...
  • Page 19: Materials

    Source container FQG61, FQG62 Materials FQG61/FQG62; order code 020, option A→  54 A0018393  8 List of materials Item Component part Material Housing 316Ti (1.4571); S235JR (1.0038) Flange 316L (1.4404) Housing ring 316L (1.4404); 304 (1.4301) Nameplate 316L (1.4404) Cover 304 (1.4301)
  • Page 20 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option B→  54 A0018394 Item Component part Material Housing 316Ti (1.4571); S235JR (1.0038) Flange 316L (1.4404) Indication plate 316L (1.4404) Rotating pin 316L (1.4404) "AUS/OFF" sign 304 (1.4301) "Source" nameplate 304 (1.4301) Rotary element 316L (1.4404)
  • Page 21 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option C→  54 A0018395 Item Component part Material Housing 316Ti (1.4571); S235JR (1.0038) Flange 316L (1.4404) Padlock: body Brass Padlock: shackle Hardened steel Indication plate 316L (1.4404) "AUS/OFF" sign 304 (1.4301) Rotary element 316L (1.4404)
  • Page 22 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option D→  54 A0018396 Item Component part Material Housing 316Ti (1.4571); S235JR (1.0038) Flange 316L (1.4404) Indication plate 316L (1.4404) 304 (1.4301) Cable 316 (1.4401) "AUS/OFF" sign 304 (1.4301) Swivel bracket 316L (1.4404)
  • Page 23 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option K, L, M or N→  54 180° A0018397 Item Component part Material Housing 316Ti (1.4571); S235JR (1.0038) Flange 316L (1.4404) Ring eyelet C15; A4 "CAUTION! Radioactive" warning sign Acrylate foil "Caution radiation"...
  • Page 24: Safety Equipment

    Source container FQG61, FQG62 Item Component part Material "Source" nameplate 304 (1.4301) Muffler G1/8 Check valve G1/8 Terminal housing nameplate (non Ex/EX) Laser foil Indication plate 316L (1.4404) Pneumatic drive Cast aluminum Item Component part Varnish Housing, flange PUR 2K texture paint RAL1003 "Australia"...
  • Page 25: Environment

    Source container FQG61, FQG62 Environment Ambient temperature range Version Ambient temperature range Manual switch-ON/switch-OFF –40 to +200 °C (–40 to +392 °F) Pneumatic switch-ON/switch-OFF –20 to +80 °C (–4 to 176 °F) If an RFID tag is used, temperature range restrictions must be taken into consideration. See...
  • Page 26: Identification

    Source container FQG61, FQG62 Identification Nameplates FQG61/FQG62; order code 020, option A→  54 Made in Germany, D-79689 Maulburg Ident-No.: Ser.No.: Order Code: Gamma Source Order Code: Radionuclide: Source No.: Activity: Date: Caution radioactive material A0018398 Nameplate of source container...
  • Page 27 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option B, C or D→  54 Made in Germany, D-79689 Maulburg Made in Germany, D-79689 Maulburg Gamma Source Order Code: FSG61-XXOO Ident no.: Ser.-No.: OOOX XX00 Ser. no.: Caution adioactive Material...
  • Page 28 Source container FQG61, FQG62 FQG61/FQG62; order code 020, option K, L, M or N→  54 Nameplate of source container Made in Germany, D-79689 Maulburg Ident no.: Ser. no.: Order code: A0034014  9 Nameplate of source container ID number of source container...
  • Page 29 250003210-- A0034016  11 Nameplate of radiation source Internal Endress+Hauser order code for the radiation source Internal Endress+Hauser serial number for the radiation source Wording "Caution Radioactive Material", if required "Cs137" or "Co60" Serial number of the source capsule (according to certificate)
  • Page 30: Rfid Tag

    Source container FQG61, FQG62 Terminal housing nameplate, non-Ex, only for option K, M Made in Germany, D-79689 Maulburg A0034018  13 Terminal housing nameplate, non-Ex, only for option K, M Maximum pressure Temperature information Degree of protection NAMUR information Circuit diagram ON...
  • Page 31 RFID TAG The RFID TAGs of the radiation source (FSG60, FSG61) and the radiation source container (FQG61, FQG62) are identical in appearance. The only difference is the data they contain and their location on the device. For additional information see:...
  • Page 32: Installation

    Source container FQG61, FQG62 Installation Incoming acceptance The radiation source container serves as Type-A packaging (IATA rules) for the radiation source. For transportation purposes, it is protected by foam packing in a box. Package dimensions: • Without pneumatic actuator: 380 x 380 x 450 mm (15 x 15 x 17.7 in) •...
  • Page 33: Mounting Instructions

    Mounting and removal is only permitted with the switch in the "OFF" position, secured by the padlock. ‣ Take the weight of the radiation source container into account when mounting: FQG61: 40 to 50 kg (88.2 to 110.25 lbs), FQG62: 87 to 97 kg (191.84 to 213.89 lbs) Endress+Hauser...
  • Page 34: Orientation For Level Measurement

    Source container FQG61, FQG62 ‣ Orientation for level For continuous level measurement, the source container must be mounted at the height of, or slightly above, the maximum level. The radiation must be aligned exactly with the detector measurement mounted on the opposite side. The source container and detector should be mounted as close as possible to the product vessel to avoid control zones.
  • Page 35: Orientation For Point Level Detection

    Source container FQG61, FQG62 ‣ Two or more source containers are generally used for large measuring ranges. The use of several sources can be necessary not only due to large measuring ranges but also for accuracy reasons.  A0018403 FQG61, FQG62 FMG60 ‣...
  • Page 36: Orientation For Density Measurement

    Source container FQG61, FQG62 ‣ Orientation for density The most constant conditions for density measurement in pipes are achieved if the unit is mounted on vertical pipelines and the feed direction is from bottom to top. If only horizontal measurement pipes are accessible, the path of the beam should also be arranged horizontally to reduce the influence of air bubbles and buildup.
  • Page 37: Mounting Device (Supplied By Customer)

    176.0 mm (6.93 in) 180.0 mm (7.09 in) 190.5 mm (7.5 in) 18.0 mm (0.71 in) 19.1 mm (0.75 in) The mounting flange of FQG61 and FQG62 is compatible with: • DN 100 PN16 • ANSI 4" 150lbs Toothed lock washers CAUTION Safety instructions ‣...
  • Page 38: Tightening Torque For Mounting Screws

    Source container FQG61, FQG62 ‣ The source container must be integrated in the potential equalization system of the plant. In order to ensure electrical contact between the radiation source container and the mounting support, the supplied toothed lock washers must be used on two of the flange screws as indicated in the adjacent graphic.
  • Page 39: Connecting The Pneumatic Actuator

    Source container FQG61, FQG62 Connecting the pneumatic actuator NOTICE This section only applies to source containers with a pneumatic actuator. (In the product structure: feature 020, versions K, L, M or N) ‣ CAUTION The pneumatic actuator may only be put into operation once the radiation source container has been mounted ‣...
  • Page 40: Commissioning

    Source container FQG61, FQG62 Proximity switches Type model: Pepperl+Fuchs 181094-NCB2-12GM35-NO-10M Cable entries Suitable cable diameter: 5 to 10 mm (0.2 to 0.39 in) Potential equalization Terminal on the cover→  39 Connection data • Nominal voltage: 8V • Current consumption •...
  • Page 41: Technical Data Of The Pneumatic Actuator

    Source container FQG61, FQG62 A0018414  20 Switch state CAUTION Risk of injury ‣ Do not touch the indicator window when the drive is pressurized. Technical data of the • Swivel range: 180° pneumatic actuator • Compressed air connection: G1/8 •...
  • Page 42: Operation

    Source container FQG61, FQG62 Operation FQG61/FQG62; order code Switching ON the radiation 020, option A CAUTION Safety instructions for switching ON the radiation ‣ Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the product vessel).
  • Page 43: Fqg61/Fqg62; Order Code 020, Option B

    Source container FQG61, FQG62 Only for source containers with order code 670 "Additional function", option WA "Density measurement > Fixation ON": screw in setscrew using Allen key. The current switching state is indicated by the arrow (B) ("EIN-ON" or "AUS-OFF").
  • Page 44 Source container FQG61, FQG62 Remove the locking bolt  A0018416 Turn the swivel bracket 180° in the counterclockwise direction. The current switch state is indicated by the sign that is visible ("EIN - ON" or "AUS - OFF"). The other sign is covered by the swivel bracket.
  • Page 45: Fqg61/Fqg62; Order Code 020, Option C

    Source container FQG61, FQG62 FQG61/FQG62; order code Switching ON the radiation 020, option C CAUTION Safety instructions for switching ON the radiation ‣ Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the product vessel).
  • Page 46: Fqg61/Fqg62; Order Code 020, Option D

    Source container FQG61, FQG62 Tighten the locking screw (optional)  A0018421 Switching OFF the radiation To switch the radiation OFF, perform the steps above in reverse order. FQG61/FQG62; order code Switching ON the radiation 020, option D CAUTION Safety instructions for switching ON the radiation ‣...
  • Page 47 Source container FQG61, FQG62 Turn the swivel bracket 180° in the counterclockwise direction. The current switch state is indicated by the sign that is visible ("EIN - ON" or "AUS - OFF"). The other sign is covered by the swivel bracket.
  • Page 48: Maintenance And Inspection

    Source containers with manual ON/OFF switching mechanism mechanism Release the locking bolt (FQG61/FQG62; order code 020, option B) or remove the padlock (if present) as described in the "Operation" section. Move the source insert several times from ON to OFF and from OFF to ON as described in the "Operation"...
  • Page 49: Routine Leak Testing

    Unless otherwise instructed, perform the leak test as follows: A0018425 FQG61/FQG62; order code 020, option A FQG61/FQG62; order code 020, option B FQG61/FQG62; order code 020, option C FQG61/FQG62; order code 020, option D FQG61/FQG62; order code 020, option K, L, M or N Endress+Hauser...
  • Page 50 Take a swab sample at the following points at least:  FQG61/FQG62; order code 020, option A, B, C, D: along the groove between the source insert and the housing FQG61/FQG62;...
  • Page 51: What To Do In An Emergency

    (e.g. steel or lead) or mount a blind flange in front of the emission channel. Individuals should at all times be behind the source container, not in front of the radiation emission channel (flange of FQG61/FQG62). The lifting eye on the housing facilitates safe handling.
  • Page 52: Procedures After Termination Of The Application

    Endress+Hauser. Other countries Contact your Endress+Hauser sales center or the appropriate authority to find a way of returning the radiation source in your country. If it is not possible to return the device in your country, the next steps to be taken must be agreed with the Endress+Hauser sales center/representative concerned.
  • Page 53 Source container FQG61, FQG62 Pre-return inspection Company Name Address Name of inspector and role Container FQG6_-_____________ Radiation source Isotope   Source serial number Nominal activity (MBq / GBq) Date of manufacture Checks Enter yes or no Wipe test report not older than 3 months is...
  • Page 54: Ordering Information

    Ordering Information Detailed ordering information is available as follows: • In the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) → Device support (right-hand column): Configure the selected product →...
  • Page 55: Accessories

    Ordering information Detailed ordering information is available as follows: • For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) →...
  • Page 56: Measurement Section Fhg62

    Ordering information Detailed ordering information is available as follows: • For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) →...