The radiation is emitted almost unattenuated in one direction only, and is damped in all other directions. FQG61 and FQG62 differ from each other in terms of size and screening effect.
Source container FQG61, FQG62 About this document Symbols used Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
Page 5
Comment TI00439F/00 • Technical Information for gamma radiation source FSG60/FSG61 • Returning source containers • Type A packaging Source container FQG60, FQG61, FQG62, FQG63, FQG66 Documentation Comment SD00297F/00 Instructions for loading and changing the source. Label Clamping device FHG61 Documentation...
Page 6
Source container FQG61, FQG62 Gammapilot M FMG60 Documentation Comment TI00363F/00 Technical Information for Gammapilot M FMG60 BA00236F/00 Operating Instructions for Gammapilot M FMG60 (HART) BA00329F/00 Operating Instructions for Gammapilot M FMG60 (PROFIBUS PA) BA00330F/00 Operating Instructions for Gammapilot M FMG60...
Safety instructions Designated use The FQG61 and FQG62 source containers described in this document contain the radioactive source, which is used for radiometric measurement of level, interface and density. They screen the radiation from the surrounding environment and allow it to be emitted almost unattenuated in the direction of measurement only.
Distance Screening Ensure the best possible screening between the radiation source and yourself and all other persons. Source containers (e.g. FQG60, FQG61/FQG62, FQG63, FQG66) and all high-density materials (lead, iron, concrete etc.) can be used for effective screening purposes. Time The time spent in the exposed area should be kept to a minimum.
The limit values for the control zone are specified in the current Radiation Protection Ordinance applicable for your area. The Endress+Hauser sales organization will be pleased to provide further information on radiation protection and regulations in other countries.
Function Function of the radiation source container In the FQG61/FQG62 source container, the radioactive source is surrounded by a steel casing filled with lead which screens off gamma radiation. The radiation can be emitted, almost unattenuated, in one direction only through a channel (focused narrow beam path). This radiation is used for radiometric measurement.
Page 12
580 "Test, Certificate". For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components →...
According to feature 240 of the product structure: • 5° • 20° • 40° Width of emission • FQG61: 10 mm (0.39 in) • FQG62: 12 mm (0.47 in) Attenuation of the Approx. 0.3 half-value layers (F = 1.2) useful beam...
Source container FQG61, FQG62 Environment Ambient temperature range Version Ambient temperature range Manual switch-ON/switch-OFF –40 to +200 °C (–40 to +392 °F) Pneumatic switch-ON/switch-OFF –20 to +80 °C (–4 to 176 °F) If an RFID tag is used, temperature range restrictions must be taken into consideration. See...
Source container FQG61, FQG62 Identification Nameplates FQG61/FQG62; order code 020, option A→ 54 Made in Germany, D-79689 Maulburg Ident-No.: Ser.No.: Order Code: Gamma Source Order Code: Radionuclide: Source No.: Activity: Date: Caution radioactive material A0018398 Nameplate of source container...
Page 27
Source container FQG61, FQG62 FQG61/FQG62; order code 020, option B, C or D→ 54 Made in Germany, D-79689 Maulburg Made in Germany, D-79689 Maulburg Gamma Source Order Code: FSG61-XXOO Ident no.: Ser.-No.: OOOX XX00 Ser. no.: Caution adioactive Material...
Page 28
Source container FQG61, FQG62 FQG61/FQG62; order code 020, option K, L, M or N→ 54 Nameplate of source container Made in Germany, D-79689 Maulburg Ident no.: Ser. no.: Order code: A0034014 9 Nameplate of source container ID number of source container...
Page 29
250003210-- A0034016 11 Nameplate of radiation source Internal Endress+Hauser order code for the radiation source Internal Endress+Hauser serial number for the radiation source Wording "Caution Radioactive Material", if required "Cs137" or "Co60" Serial number of the source capsule (according to certificate)
Source container FQG61, FQG62 Terminal housing nameplate, non-Ex, only for option K, M Made in Germany, D-79689 Maulburg A0034018 13 Terminal housing nameplate, non-Ex, only for option K, M Maximum pressure Temperature information Degree of protection NAMUR information Circuit diagram ON...
Page 31
RFID TAG The RFID TAGs of the radiation source (FSG60, FSG61) and the radiation source container (FQG61, FQG62) are identical in appearance. The only difference is the data they contain and their location on the device. For additional information see:...
Source container FQG61, FQG62 Installation Incoming acceptance The radiation source container serves as Type-A packaging (IATA rules) for the radiation source. For transportation purposes, it is protected by foam packing in a box. Package dimensions: • Without pneumatic actuator: 380 x 380 x 450 mm (15 x 15 x 17.7 in) •...
Mounting and removal is only permitted with the switch in the "OFF" position, secured by the padlock. ‣ Take the weight of the radiation source container into account when mounting: FQG61: 40 to 50 kg (88.2 to 110.25 lbs), FQG62: 87 to 97 kg (191.84 to 213.89 lbs) Endress+Hauser...
Source container FQG61, FQG62 ‣ Orientation for level For continuous level measurement, the source container must be mounted at the height of, or slightly above, the maximum level. The radiation must be aligned exactly with the detector measurement mounted on the opposite side. The source container and detector should be mounted as close as possible to the product vessel to avoid control zones.
Source container FQG61, FQG62 ‣ Two or more source containers are generally used for large measuring ranges. The use of several sources can be necessary not only due to large measuring ranges but also for accuracy reasons. A0018403 FQG61, FQG62 FMG60 ‣...
Source container FQG61, FQG62 ‣ Orientation for density The most constant conditions for density measurement in pipes are achieved if the unit is mounted on vertical pipelines and the feed direction is from bottom to top. If only horizontal measurement pipes are accessible, the path of the beam should also be arranged horizontally to reduce the influence of air bubbles and buildup.
176.0 mm (6.93 in) 180.0 mm (7.09 in) 190.5 mm (7.5 in) 18.0 mm (0.71 in) 19.1 mm (0.75 in) The mounting flange of FQG61 and FQG62 is compatible with: • DN 100 PN16 • ANSI 4" 150lbs Toothed lock washers CAUTION Safety instructions ‣...
Source container FQG61, FQG62 ‣ The source container must be integrated in the potential equalization system of the plant. In order to ensure electrical contact between the radiation source container and the mounting support, the supplied toothed lock washers must be used on two of the flange screws as indicated in the adjacent graphic.
Source container FQG61, FQG62 Connecting the pneumatic actuator NOTICE This section only applies to source containers with a pneumatic actuator. (In the product structure: feature 020, versions K, L, M or N) ‣ CAUTION The pneumatic actuator may only be put into operation once the radiation source container has been mounted ‣...
Source container FQG61, FQG62 A0018414 20 Switch state CAUTION Risk of injury ‣ Do not touch the indicator window when the drive is pressurized. Technical data of the • Swivel range: 180° pneumatic actuator • Compressed air connection: G1/8 •...
Source container FQG61, FQG62 Operation FQG61/FQG62; order code Switching ON the radiation 020, option A CAUTION Safety instructions for switching ON the radiation ‣ Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the product vessel).
Source container FQG61, FQG62 Only for source containers with order code 670 "Additional function", option WA "Density measurement > Fixation ON": screw in setscrew using Allen key. The current switching state is indicated by the arrow (B) ("EIN-ON" or "AUS-OFF").
Page 44
Source container FQG61, FQG62 Remove the locking bolt A0018416 Turn the swivel bracket 180° in the counterclockwise direction. The current switch state is indicated by the sign that is visible ("EIN - ON" or "AUS - OFF"). The other sign is covered by the swivel bracket.
Source container FQG61, FQG62 FQG61/FQG62; order code Switching ON the radiation 020, option C CAUTION Safety instructions for switching ON the radiation ‣ Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the product vessel).
Source container FQG61, FQG62 Tighten the locking screw (optional) A0018421 Switching OFF the radiation To switch the radiation OFF, perform the steps above in reverse order. FQG61/FQG62; order code Switching ON the radiation 020, option D CAUTION Safety instructions for switching ON the radiation ‣...
Page 47
Source container FQG61, FQG62 Turn the swivel bracket 180° in the counterclockwise direction. The current switch state is indicated by the sign that is visible ("EIN - ON" or "AUS - OFF"). The other sign is covered by the swivel bracket.
Source containers with manual ON/OFF switching mechanism mechanism Release the locking bolt (FQG61/FQG62; order code 020, option B) or remove the padlock (if present) as described in the "Operation" section. Move the source insert several times from ON to OFF and from OFF to ON as described in the "Operation"...
Unless otherwise instructed, perform the leak test as follows: A0018425 FQG61/FQG62; order code 020, option A FQG61/FQG62; order code 020, option B FQG61/FQG62; order code 020, option C FQG61/FQG62; order code 020, option D FQG61/FQG62; order code 020, option K, L, M or N Endress+Hauser...
Page 50
Take a swab sample at the following points at least: FQG61/FQG62; order code 020, option A, B, C, D: along the groove between the source insert and the housing FQG61/FQG62;...
(e.g. steel or lead) or mount a blind flange in front of the emission channel. Individuals should at all times be behind the source container, not in front of the radiation emission channel (flange of FQG61/FQG62). The lifting eye on the housing facilitates safe handling.
Endress+Hauser. Other countries Contact your Endress+Hauser sales center or the appropriate authority to find a way of returning the radiation source in your country. If it is not possible to return the device in your country, the next steps to be taken must be agreed with the Endress+Hauser sales center/representative concerned.
Page 53
Source container FQG61, FQG62 Pre-return inspection Company Name Address Name of inspector and role Container FQG6_-_____________ Radiation source Isotope Source serial number Nominal activity (MBq / GBq) Date of manufacture Checks Enter yes or no Wipe test report not older than 3 months is...
Ordering Information Detailed ordering information is available as follows: • In the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) → Device support (right-hand column): Configure the selected product →...
Ordering information Detailed ordering information is available as follows: • For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) →...
Ordering information Detailed ordering information is available as follows: • For the allocation of the device version, see the Product Configurator on the Endress+Hauser website: www.endress.com → Select your country → Products → Select measuring technology, software or components → Select the product (picklists: measurement method, product family etc.) →...