Fluke Dry-Well 9103 User Manual page 26

Hide thumbs Also See for Dry-Well 9103:
Table of Contents

Advertisement

9103
User's Guide
Command Description
Configuration Menu
Operating Parameters Menu
Read High Limit
Set High Limit
Serial Interface Menu
Read serial sample setting
Set serial sampling setting to n seconds
Set serial duplex mode:
Set serial duplex mode to full
Set serial duplex mode to half
Set serial linefeed mode:
Set serial linefeed mode to on
Set serial linefeed mode to off
Calibration Menu
Read R0 calibration parameter
Set R0 calibration parameter to n
Read ALPHA calibration parameter
Set ALPHA calibration parameter to n
Read DELTA calibration parameter
Set DELTA calibration parameter
Read BETA calibration parameter
Set BETA calibration parameter
Functions not on menu
Read firmware version number
Read structure of all commands
Read all operating parameters
Legend:
[] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a carriage return
and linefeed. Then the value is returned as indicated in the RETURNED column.
Command
Command
Format
Example
hl
hl
hl=n
hl=90
sa[mple]
sa
sa[mple]=n
sa=0
du[plex]=f[ull]/
h[alf]
du[plex]=f[ull]
du=f
du[plex]=h[alf]
du=h
lf[eed]=on/of[f]
lf[eed]=on
lf=on
lf[eed]=of[f]
lf=of
r[0]
r
r[0]=n
r=100.324
al[pha]
al
al[pha]=n
al=0.0038433
de[lta]
de
de[lta]=n
de=1.3742
be[ta]
be
be[ta]=n
be=0.342
*ver[sion]
*ver
h[elp]
h
all
all
Returned
Returned
Example
hl: 999
hl: 140
sa: 9
sa: 1
r0: 999.999
r0: 100.578
al: 9.9999999
al: 0.0038573
de: 9.99999
de: 1.507
be:9.999
be:0.342
ver.9999,9.99
ver.9103,2.00
list of commands
list of parameters
Acceptable
Values
0-140
0 to 999
FULL or HALF
ON or OFF
90 to 110
.002 to .005
0–3.0
-100.0 to 100.0
22

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dry-well 9140

Table of Contents