Endress+Hauser PROline promag 23 Operating Instructions Manual

Endress+Hauser PROline promag 23 Operating Instructions Manual

Electromagnetic flow measuring system two-wire, loop-powered
Table of Contents

Advertisement

Quick Links

BA 045D/06/en/02.03
50097235
Valid as of software version:
V 2.00.01 (amplifier)
V 2.0X.XX (communication)
PROline promag 23
Electromagnetic
Flow Measuring System
Two-wire, loop-powered
Operating Instructions

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser PROline promag 23

  • Page 1 PROline promag 23 BA 045D/06/en/02.03 50097235 Electromagnetic Valid as of software version: V 2.00.01 (amplifier) V 2.0X.XX (communication) Flow Measuring System Two-wire, loop-powered Operating Instructions...
  • Page 2 Returning devices If you return a measuring device to Endress+Hauser for repair or calibration, you must complete the “Declaration of contamination” form and enclose it with the device. You will find a preprinted blank of this form at the back of this manual.
  • Page 3: Table Of Contents

    PROline Promag 23 Contents Contents Safety instructions Error messages ......45 ....
  • Page 4 Contents PROline Promag 23 Endress+Hauser...
  • Page 5: Safety Instructions

    Strict compliance with the instructions in the Operating Manual is mandatory. x Endress+Hauser will be happy to assist in clarifying the chemical resistance proper- ties of parts wetted by special mediums, including mediums used for cleaning. x If welding work is performed on the piping system, do not ground the welding appli- ance through the Promag flowmeter.
  • Page 6: Return

    Endress+Hauser: x Always enclose a duly completed “Declaration of contamination” form. Only then can Endress+Hauser transport, examine and repair a returned device. x Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/155/EEC.
  • Page 7: Identification

    PROline Promag 23 2 Identification Identification Device designation The “Promag 23” flow measuring system consists of the following components: x Promag 23 transmitter x Promag P or Promag H sensor Transmitter and sensor form a single mechanical unit. 2.1.1 Nameplate of the transmitter...
  • Page 8: Nameplate Of The Sensor

    2 Identification PROline Promag 23 2.1.2 Nameplate of the sensor ENDRESS+HAUSER PROMAG XXXXX-XXXXXXXXXXXX Order Code: 12345678901 Ser.No.: ABCDEFGHJKLMNPQRST TAG No.: 0.8328 / –5 K-factor: DN100 DIN PN 16 150°C TMmax.: PFA / 1.4435 Materials: EPD/MSU R/B 0.5% CAL IP67/NEMA/Type4X –20°C (–4°F) < Tamb < +60°C+140°F Pat.
  • Page 9: Ce Mark, Declaration Of Conformity

    Laboratory Procedures” and with the requirements of EN 61326/A1. The measuring system described in this Operating Manual is therefore in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms suc- cessful testing of the device by affixing to it the CE mark.
  • Page 10 2 Identification PROline Promag 23 Endress+Hauser...
  • Page 11: Installation

    PROline Promag 23 3 Installation Installation Incoming acceptance, transport and storage 3.1.1 Incoming acceptance x Check the packaging and the contents for damage. x Check the shipment, make sure nothing is missing and that the scope of supply mat- ches your order.
  • Page 12: Storage

    3 Installation PROline Promag 23 3.1.3 Storage x Pack the measuring device in such a way as to protect it reliably against impact for storage (and transportation). The original packaging provides optimum protection. x The permissible storage temperature is  10...+50 °C (preferably +20 °C).
  • Page 13: Installation Conditions

    PROline Promag 23 3 Installation Installation conditions 3.2.1 Dimensions Dimensions and the fitting lengths of the transmitter and sensor are on Page 85 ff. 3.2.2 Mounting location Correct measuring is possible only if the pipe is full. Avoid the following locations: x Highest point in the run.
  • Page 14 3 Installation PROline Promag 23 Partially filled pipes Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection function (see Page 54) offers additional protection by detecting empty or partially filled pipes. " Caution: Risk of solids accumulating. Do not install the sensor at the lowest point in the drain.
  • Page 15: Orientation

    PROline Promag 23 3 Installation 3.2.3 Orientation An optimum orientation position helps avoid gas and air accumulations and deposits in the measuring tube. Promag, nevertheless, supplies a range of functions and accesso- ries for correct measuring of problematic fluids: Empty Pipe Detection (EPD) ensures the detection of partially filled measuring tubes, e.g.
  • Page 16: Inlet And Outlet Runs

    3 Installation PROline Promag 23 3.2.4 Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy:...
  • Page 17: Adapters

    PROline Promag 23 3 Installation 3.2.6 Adapters Suitable adapters to (E) DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
  • Page 18: Nominal Diameter And Flow Rate

    3 Installation PROline Promag 23 3.2.7 Nominal diameter and flow rate The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity of flow is 2…3 m/s. The velocity of flow (v), moreover, has to be matched to the physical properties of the fluid: x v <...
  • Page 19 PROline Promag 23 3 Installation Promag H Flow rate characteristic values - Promag H (SI units) Nominal Recommended Factory settings diameter flow rate min./max. full scale value Full scale value Pulse value Low flow cut off [mm] inch] (v ~ 0.3 or 10 m/s) (v ~ 2.5 m/s)
  • Page 20: Installation

    3 Installation PROline Promag 23 Installation 3.3.1 Installing the Promag P sensor " Caution: x The protective covers mounted on the two sensor flanges guard the Teflon (PTFE) lining, which is turned over the flanges. Consequently, do not remove these covers until immediately before the sensor is installed in the pipe.
  • Page 21 PROline Promag 23 3 Installation Assembly with ground disks (DN 25…200) Depending on the application, e.g. with lined or ungrounded pipes (see Page 35 ff.), it may be necessary to mount ground disks between the sensor and the pipe flange for the potential equalisation.
  • Page 22 3 Installation PROline Promag 23 Installing the high temperature version (with PFA lining) The high temperature version has a housing support for the thermal separation of sensor and transmitter. The high temperature version is always used for applications in which high ambient temperatures are encountered in conjunction with high fluid temperatures.
  • Page 23 PROline Promag 23 3 Installation Tightening torques for threaded fasteners (Promag P) Note the following points: x The tightening torques listed below are for lubricated threads only. x Always tighten threaded fasteners uniformly and in diagonally opposite sequence. x Overtightening the fasteners will deform the sealing faces or damage the seals.
  • Page 24 3 Installation PROline Promag 23 Promag P Threaded Max. tightening torque [Nm] Nominal diameter Pressure rating fasteners [mm] PTFE 4 x M 16 4 x M 16 4 x M 16 4 x M 16 8 x M 16 4 x M 16...
  • Page 25: Installing The Promag H Sensor

    PROline Promag 23 3 Installation 3.3.2 Installing the Promag H sensor The Promag H is supplied to order, with or without pre-installed process connections. Pre-installed process connections are secured to the sensor with hex-head threaded fasteners. " Caution: x If you intend using your own process connections, make up the process adapters as specified on Page 90 ff.
  • Page 26 3 Installation PROline Promag 23 Usage and assembly of ground rings (DN 2…25) In case the process connections are made of plastic (e.g. flanges or adhesive fittings), the potential between the sensor and the fluid must be equalised using additional ground rings.
  • Page 27: Turning The Transmitter Housing

    PROline Promag 23 3 Installation Welding the sensor into the piping (welding adapters) " Caution: Risk of destroying the measuring electronics. Make sure that the welding machine is not grounded via the sensor or the transmitter. Tack-weld the Promag H sensor into the pipe. A suitable welding jig can be ordered separately from E+H as an accessory (see Page 59).
  • Page 28: Turning The Local Display

    3 Installation PROline Promag 23 3.3.4 Turning the local display Remove the cover of the electronics compartment. Press in the latches at the side and turn the display module to the desired position (45° steps in either direction). Screw the cover of the electronics compartment firmly onto the transmitter housing.
  • Page 29: Post-Installation Check

    PROline Promag 23 3 Installation Post-installation check Perform the following checks after installing the measuring device in the pipe: Device condition and specifications Notes Is the device damaged (visual inspection)? Does the device correspond to specifications at the measuring point, see Page 75 ff.
  • Page 30 3 Installation PROline Promag 23 Endress+Hauser...
  • Page 31: Wiring

    PROline Promag 23 4 Wiring Wiring Warning: When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to this Operating Manual. Please do not hesitate to contact your E+H representative if you have any questions. Connecting the measuring unit 4.1.1...
  • Page 32: Terminal Assignment

    4 Wiring PROline Promag 23 4.1.2 Terminal assignment Outputs Terminal No. 1(+) / 2(  3(+) / 4(  Order variant 23***-***********W Current output HART 23***-***********A Current output HART Frequency output Mandatory: A common connecting cable carries supply voltage and measuring output signal: Current output (passive) galvanically isolated: 12...30 V DC (Ex i: 13,9...30 V DC), 4...20 mA...
  • Page 33: Load

    PROline Promag 23 4 Wiring 4.1.3 Load Non Ex area: V > @ U V > @ V > @ 12 V > @ – – > @ ------------------------------------ - ------------------------------------ - > @ > @ 0 022 A Ex area (Ex i): V >...
  • Page 34: Hart Connection

    4 Wiring PROline Promag 23 4.1.4 HART connection Users have the following connection options at their disposal: x Direct connection to transmitter by means of terminals 1 / 2 x Connection by means of the 4…20 mA circuit Note: x The measuring loop's minimum load must be at least 250 : .
  • Page 35: Potential Equalisation

    PROline Promag 23 4 Wiring Potential equalisation 4.2.1 Standard case Perfect measurement is only ensured when the medium and the sensor have the same electrical potential. Most Promag sensors have a standard installed reference electrode which guarantees the required potential matching. This usually means that additional potential matching measures are unnecessary.
  • Page 36: Special Cases

    4 Wiring PROline Promag 23 4.2.2 Special cases Metal, ungrounded piping In order to prevent outside influences on measurement, it is advisable to use ground cables to connect each sensor flange to its corresponding pipe flange and ground the flanges. Connect the transmitter or sensor connection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose (Fig.
  • Page 37 PROline Promag 23 4 Wiring Plastic pipes and isolating lined pipes Normally, potential is matched using the reference electrodes in the measuring tube. However, in exceptional cases it is possible that, due to the grounding plan of a system, large matching currents flow over the reference electrodes. This can lead to destruction of the sensor, e.g.
  • Page 38: Degree Of Protection

    Do not remove the grommet from the cable entry. Fig. 28: Installation instructions, cable entries " Caution: Do not loosen the threaded fasteners of the Promag sensor housing, as otherwise the degree of protection guaranteed by Endress+Hauser no longer applies. Endress+Hauser...
  • Page 39: Post-Connection Check

    PROline Promag 23 4 Wiring Post-connection check Perform the following checks after completing electrical installation of the measuring device: Device condition and specifications Notes Are cables or the device damaged (visual inspection)? Electrical connection Notes Does the supply voltage match the specifications on the nameplate? 12...30 V DC...
  • Page 40 4 Wiring PROline Promag 23 Endress+Hauser...
  • Page 41: Operation

    PROline Promag 23 5 Operation Operation Display and operating elements The local display enables you to read all important parameters directly at the measuring point or to configure the device via the function matrix. +24.502 +24.502 +1863.97 +1863.97 –50 –50 Fig.
  • Page 42 5 Operation PROline Promag 23 Display (operation mode) The display area consists of three lines in all; this is where measured values are dis- played, and/or status variables (direction of flow, partially filled pipe, bargraph, etc.). You can change the assignment of display lines to variables at will in order to customize the display to suit your needs and preferences ( o see the “Description of Device...
  • Page 43: Brief Operating Instruction To The Function Matrix

    PROline Promag 23 5 Operation Brief operating instruction to the function matrix Note: x See the general notes on Page 44. x Function descriptions o see the “Description of Device Functions” manual HOME position o F o Enter the function matrix Select a block (e.g.
  • Page 44: General Notes

    There is no need to change these parameters under normal circumstances and conse- quently, they are protected by a special code known only to the E+H service organisa- tion. Please contact Endress+Hauser if you have any questions. Endress+Hauser...
  • Page 45: Disabling The Programming Mode

    PROline Promag 23 5 Operation 5.2.3 Disabling the programming mode Programming mode is disabled if you do not press a key within 60 seconds following automatic return to the HOME position. You can also disable programming in the “ACCESS CODE” function by entering any number (other than the customer's code).
  • Page 46 5 Operation PROline Promag 23 Error message type Users have the option of weighting certain errors differently, in other words having them classed as “Fault messages” or “Notice messages”. You can define messages in this way with the aid of the function matrix see the “Description of Device Functions”...
  • Page 47: Communication (Hart)

    PROline Promag 23 5 Operation Communication (HART) In addition to local operation, the measuring device can be configured and measured values can be obtained by means of the HART protocol. Digital communication takes place using the 4–20 mA current output HART (see Page 34).
  • Page 48: Universal Hart Commands

    5 Operation PROline Promag 23 Other operating programs x Operating program “AMS” (Fisher Rosemount) x Operating program “SIMATIC PDM” (Siemens) Note: x The HART protocol requires the “4…20 mA HART” setting in the CURRENT SPAN function (current output). x HART write protection can be disabled or enabled by means of a jumper on the I/O board o Page 51.
  • Page 49 PROline Promag 23 5 Operation Command No. Command data Response data HART command / Access type (numeric data in decimal form) (numeric data in decimal form) Read the primary process none 24 bytes are sent as a response: variable as current in mA and –...
  • Page 50 5 Operation PROline Promag 23 Command No. Command data Response data HART command / Access type (numeric data in decimal form) (numeric data in decimal form) Read TAG, descriptor and none – Bytes 0-5: TAG date – Bytes 6-17: descriptor –...
  • Page 51: Switching Hart Write Protection On And Off

    PROline Promag 23 5 Operation 5.4.3 Switching HART write protection on and off A jumper on the I/O board provides the means of activating or deactivating HART write protection. Warning: Make sure that the power supply is switched off before you remove the cover of the elec- tronics compartment.
  • Page 52 5 Operation PROline Promag 23 Endress+Hauser...
  • Page 53: Commissioning

    PROline Promag 23 6 Commissioning Commissioning Function check Make sure that all final checks have been completed before you start up your measuring point: x Checklist for “Post-installation check” o Page 29 x Checklist for “Post-connection check” o Page 39 Commissioning 6.2.1...
  • Page 54: Empty-Pipe/Full-Pipe Calibration

    6 Commissioning PROline Promag 23 6.2.2 Empty-pipe/full-pipe calibration Flow cannot be measured correctly unless the measuring tube is full. This status can be monitored at all times with the Empty Pipe Detection function (EPD). The EPD function cannot be activated until after empty/full pipe calibration has been performed.
  • Page 55: Data Storage Device

    PROline Promag 23 6 Commissioning Data storage device S-DAT (Sensor-DAT) The S-DAT is an exchangeable data storage device in which all sensor relevant para- meters are stored, i.e., diameter, serial number, calibration factor, zero point. T-DAT (Transmitter-DAT) The T-DAT is an exchangeable data storage device in which all transmitter parameters and settings are stored.
  • Page 56 6 Commissioning PROline Promag 23 Endress+Hauser...
  • Page 57: Maintenance

    PROline Promag 23 7 Maintenance Maintenance The Promag 23 flow measuring system requires no special maintenance. Exterior cleaning When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the surface of the housing and the seals.
  • Page 58 7 Maintenance PROline Promag 23 Endress+Hauser...
  • Page 59: Accessories

    PROline Promag 23 8 Accessories Accessories Various accessories, which can be ordered separately from Endress+Hauser, are avai- lable for the transmitter and the sensor. The E+H service organisation can provide detailed information on the order codes of your choice. Accessory...
  • Page 60 8 Accessories PROline Promag 23 Accessory Description Ordering code DXS10  * * * * * FieldTool Configuration and service software for flow- meters in the field: – Commissioning, maintenance analysis – Configuring flowmeters – Service functions – Visualisation of process data –...
  • Page 61: Trouble-Shooting

    PROline Promag 23 9 Trouble-shooting Trouble-shooting Trouble-shooting instructions Always start trouble-shooting with the checklist below, if faults occur after start-up or during operation. The routine takes you directly to the cause of the problem and the appropriate remedial measures. Check the display 1.
  • Page 62: System Error Messages

    9 Trouble-shooting PROline Promag 23 System error messages Serious system errors are always recognised by the instrument as “Fault message”, and are shown as a lightning flash ($) on the display. Fault messages immediately affect the inputs and outputs. Simulations and positive zero return, on the other hand, are classed and displayed as notice messages.
  • Page 63 PROline Promag 23 9 Trouble-shooting Type Error message / No. Cause Remedy / spare part SENSOR SW DAT Sensor: 1. Check whether the S-DAT is # 032 Error accessing the calibration correctly plugged into the values stored in the S-DAT.
  • Page 64 9 Trouble-shooting PROline Promag 23 Type Error message / No. Cause Remedy / spare part Nr. # 2xx Error in DAT / no communication LOAD T-DAT Data backup (downloading) to 1. Check whether the T-DAT is # 205 T-DAT failed, or error when...
  • Page 65 PROline Promag 23 9 Trouble-shooting Type Error message / No. Cause Remedy / spare part PULSE RANGE Pulse output: 1. Increase the setting for pulse # 359 Pulse output frequency is out of weighting range. 2. When selecting the pulse...
  • Page 66: Process Error Messages

    9 Trouble-shooting PROline Promag 23 Process error messages Process errors can be defined as either “Fault” or “Notice” messages and can thereby be weighted differently. Determination of this is done via the function matrix (see the “Description of Device Functions” Manual).
  • Page 67: Process Errors Without Error Messages

    Returning devices to E+H The procedures on Page 6 must be carried out before you return a flow- meter requiring repair or calibration to Endress+Hauser. Always enclose a duly completed “Declaration of contamination” form with the flowmeter. You will find a preprinted form at the back of this manual.
  • Page 68 9 Trouble-shooting PROline Promag 23 Symptoms Rectification Choice T-DAT LOAD not The T-DAT is either not plugged or has no data on it. available on display. In the first case plug the T-DAT in the socket. In the second case write first data on it using the function SAVE.
  • Page 69: Response Of Outputs To Errors

    PROline Promag 23 9 Trouble-shooting Response of outputs to errors Note: The failsafe mode of totalizers, current, pulse and frequency outputs can be customized by means of various functions in the function matrix. You will find detailed information on these procedures in the “Description of Device Functions” manual.
  • Page 70 9 Trouble-shooting PROline Promag 23 Failsafe mode of outputs and totalizers Process/system error is current Positive zero return is activated Totalizer STOP Totalizer stops The totalizers are paused until the error is rectified. ACTUAL VALUE The fault is ignored. The totalizers continue to count in accordance with the current flow value.
  • Page 71: Spare Parts

    PROline Promag 23 9 Trouble-shooting Spare parts Section 9.1 contains a detailed trouble-shooting guide. The measuring device, moreover, provides additional support in the form of continuous self-diagnosis and error messages. Fault rectification can entail replacing defective components with tested spare parts.
  • Page 72: Removing And Installing Printed Circuit Boards

    – Always remove the I/O board first and reinstall it last. – When reassembling, make sure that the coil current cable (8) and the electrode signal cable (5) are correctly reinserted into their retaining clips! – Use only original Endress+Hauser parts. Endress+Hauser...
  • Page 73 PROline Promag 23 9 Trouble-shooting Fig. 35: Removing and installing printed circuit boards Housing cover Mounting plate display module (2 screws) Display module Ribbon cable (display module) Electrode signal cable (sensor) Cover of electronics compartment (1 screw) Connecting cable, printed circuit boards...
  • Page 74: Software History

    9 Trouble-shooting PROline Promag 23 Software history Software version / Changes to software Changes to documentation date Amplifier: V 1.00.00 / 05.2000 Original software. Compatible with: – FieldTool – Commuwin II (version 2.05.03 and hig- her) – HART Communicator DXR 275 (from OS 4.6) with Rev.
  • Page 75: Technical Data

    PROline Promag 23 10 Technical data Technical data 10.1 Technical data at a glance 10.1.1 Application x Measuring the flow rate of fluids in closed piping systems. x A minimum conductivity of t 50 P S/cm is required for measuring.
  • Page 76: Power Supply

    10 Technical data PROline Promag 23 x Ex i version: – Power supply, signal circuits and pulse/frequency output with “intrinsically safe” protection rating, EEx ia IIC and EEx ia IIB, only for connection to certified, intrinsi- cally safe circuits with the following maximum values: U...
  • Page 77: Performance Characteristics

    PROline Promag 23 10 Technical data 10.1.6 Performance characteristics Reference operating To DIN 19200 and VDI/VDE 2641: x Fluid temperature: +28 °C ± 2 K conditions x Ambient temperature: +22 °C ± 2 K x Warm-up time: 30 minutes Installation: x Inlet run >10 x DN...
  • Page 78 10 Technical data PROline Promag 23 Shock and Acceleration up to 2 g by analogy with IEC 68-2-6 vibration resistance (high temperature version: no data available) CIP cleaning Promag P: possible (note max. temperature) Promag H: possible (note max. temperature) SIP cleaning Promag P: possible with PFA (note max.
  • Page 79 PROline Promag 23 10 Technical data Minimum conductivity: 50 P S/cm (for fluids generally) Conductivity Limiting medium pressure Promag P range (nominal pressure) DIN 2501: PN 10 (DN 200) PN 16 (DN 65...200) PN 25 (DN 200) PN 40 (DN 25...150) ANSI B16.5:...
  • Page 80: Mechanical Construction

    10 Technical data PROline Promag 23 10.1.8 Mechanical construction Design / dimensions see Page 85 ff. Weight Weight data of Promag P in kg Nominal diameter Compact version [mm] [inch] ANSI 1" 1 1/4" – 1 1/2" 2" 10,1 10,1 2 1/2"...
  • Page 81 PROline Promag 23 10 Technical data Materials Promag P Transmitter housing: Compact housing: powder coated die-cast aluminium Sensor housing: powder coated die-cast aluminium Measuring tube: stainless steel 1.4301 or 1.4306/304L; non-stainless flange material with AI/Zn protective coating Flange: x DIN : Stainless steel 1.4571, ST37 / FE 410W B (with Al/Zn protective coating)
  • Page 82: Human Interface

    Seals in conformity with FDA (except Kalrez seals) CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. Endress+Hauser...
  • Page 83: 10.1.11 Ordering Information

    PROline Promag 23 10 Technical data Other standards and EN 60529 guidelines Degrees of protection by housing (IP code) EN 61010 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures EN 61326/A1 (IEC 1326) Electromagnetic compatibility (EMC requirements) NAMUR NE 21 Association for Standards for Control and Regulation in the Chemical Industry.
  • Page 84: Measuring-Tube Specifications

    10 Technical data PROline Promag 23 10.2 Measuring-tube specifications Promag P Pressure rating Inside diameter of Nominal diameter measuring tube ANSI ANSI with PFA with PTFE [mm] [inch] [bar] [lbs] [mm] [mm] 1" PN 40 Cl 150 PN 40 1 1/2"...
  • Page 85: Dimensions Promag 23 P

    PROline Promag 23 10 Technical data 10.3 Dimensions Promag 23 P Promag P Fig. 37: Dimensions Promag P ANSI [mm] [inch] [mm] [mm] [mm] [mm] [mm] [mm] 1" 1 1/2" 2" 3" 4" 6" 8" – The fitting length (L) is always the same, regardless of the pressure rating.
  • Page 86 10 Technical data PROline Promag 23 Promag P / high temperature version Fig. 38: Dimensions of high temperature version (Promag P) Dimensions A1, B1 = A, B of standard version plus 110 mm Endress+Hauser...
  • Page 87: Dimensions Of Ground Disks (Promag P)

    PROline Promag 23 10 Technical data 10.4 Dimensions of ground disks (Promag P) Ø B Ø A Ø D Fig. 39: Dimensions of ground disks (Promag P / DN 25…200) ANSI [mm] [inch] [mm] [mm] [mm] [mm] 1" 77.5 87.5 38.5...
  • Page 88: Dimensions Promag 23 H

    10 Technical data PROline Promag 23 10.5 Dimensions Promag 23 H Promag H / DN 2...25 Fig. 40: Dimensions Promag H / DN 2...25 PN ** ANSI [mm] [inch] [bar] [mm] [mm] [mm] [mm] 16/40 2.25 M 6x4 16/40 M 6x4...
  • Page 89 PROline Promag 23 10 Technical data Promag H / DN 40...100 Fig. 42: Dimensions Promag H / DN 40...100 ANSI [mm] [inch] [bar] [mm] [mm] [mm] [mm] [mm] [mm] [mm] 1 1/2" 35.3 M 6x4 2" 48.1 M 8x4 2 1/2"...
  • Page 90 10 Technical data PROline Promag 23 10.6 Process connections Promag H (DN 2…25) Front view of sensor Promag H / DN 2…25 (without process connections) DN 2...15 DN 25 ±0.1 ±0.1 ±0.1 ±0.1 Fig. 43: Dimensions front view of sensor DN 2…25...
  • Page 91 PROline Promag 23 10 Technical data Process connections with O-ring seals (DN 2…25) Weld nipples for ISO 2463 pipe Sensor Pipe H x B 1.4404 / 316L DN [mm] Tube [mm] [mm] [mm] [mm] 5*H**-B*********** 2…8 13.5 x 1.6 10.3 13.5...
  • Page 92 10 Technical data PROline Promag 23 Flange 20 K / JIS B2238 Sensor Pipe H x B 1.4404 / 316L DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm] 5*H**-F*********** 2…8 ND 10 60 x 42 ND 15 60 x 42...
  • Page 93 PROline Promag 23 10 Technical data External pipe thread ISO 228 / Sensor Pipe H x B DIN 2999, 1.4404 / 316L Internal thread 5*H**-K*********** DN [mm] [inch] [mm] [inch] [mm] [mm] [mm] 2…8 R 3/8" 3/8" 10.1 60 x 42 R 1/2"...
  • Page 94 10 Technical data PROline Promag 23 Process connections with aseptic gasket seal (DN 2…25 ) Weld nipple for DIN 11850 pipe Sensor Pipe H x B 1.4404 / 316L DN [mm] Tube [mm] [mm] [mm] [mm] 5*H**-U*********** 2…8 14 x 2 23.3...
  • Page 95 PROline Promag 23 10 Technical data Tri-clamp Sensor Pipe H x B 1.4404 / 316L DN [mm] Tri-Clamp [mm] [mm] [mm] [mm] 5*H**-1*********** 2…8 Tube 12.7 x 1.65 25.0 28.5 60 x 42 (ODT 1/2") Tube 19.1 x 1.65 15.8 25.0...
  • Page 96 10 Technical data PROline Promag 23 Coupling SMS 1145 Sensor Pipe H x B 1.4404 / 316L DN [mm] Coupling [inch] [mm] [mm] [mm] [mm] 5*H**-5*********** 1" (25 ANSI) 1" 22.1 Rd 40 x 1/6" 30.8 70 x 52 – Fitting length = (2 x L) + 86 mm –...
  • Page 97 PROline Promag 23 10 Technical data Process connections orderable only as accessories (with aseptic gasket seal) Tri-Clamp Sensor Pipe H x B 1.4404 / 316L DN [mm] Tri-Clamp [mm] [mm] [mm] [mm] DKH**-HF*** Tube 25.4 x 1.5 22.1 50.4 28.5 60 x 42 (ODT;...
  • Page 98: Process Connections Of

    10 Technical data PROline Promag 23 10.7 Process connections of Promag H (DN 40…100) Front view of sensor Promag H / DN 40…100 (without process connection) Fig. 44: Dimensions front view of sensor DN 40…100 [mm] [mm] [mm] [mm] [mm]...
  • Page 99 PROline Promag 23 10 Technical data Process connections with gasket seal (DN 40…100) Weld nipples for DIN 11850 pipe 1.4404 / 316L [mm] [mm] [mm] [mm] [mm] [mm] [mm] 5*H**-U*********** 38.0 71.0 50.0 83.5 66.0 100.0 81.0 121.0 100.0 141.5 –...
  • Page 100 10 Technical data PROline Promag 23 Tri-Clamp 1.4404 / 316L [mm] [inch] [mm] [mm] [mm] [mm] [mm] 5*H**-1*********** 1 1/2" 34.8 50.4 68.6 71.0 2" 47.5 63.9 68.6 83.5 60.2 77.4 68.6 100.0 3" 72.9 90.9 68.6 121.0 4" 97.4 118.9...
  • Page 101 PROline Promag 23 10 Technical data Coupling SMS 1145 1.4404 / 316L [mm] [mm] [mm] [mm] [mm] [mm] 5*H**-5*********** 35.5 Rd 60 x 1/6" 71.0 48.5 Rd 70 x 1/6" 83.5 60.5 Rd 85 x 1/6" 100.0 72.0 Rd 98 x 1/6"...
  • Page 102 10 Technical data PROline Promag 23 Endress+Hauser...
  • Page 103: Index

    PROline Promag 23 Index Accessories ....... 59 Electrical connection Adapters (installation of sensors) ....17 Degree of protection .
  • Page 104 PROline Promag 23 Measuring range ......75 Measuring system ......75...
  • Page 105 PROline Promag 23 Registered trademarks ..... . . 9 Weights ........80 Repair .
  • Page 106 PROline Promag 23 Endress+Hauser...
  • Page 107 Declaration of contamination Dear customer, Because of legal determinations and for the safety of our employes and operating equipment we need this “Declaration of contamination” with your signature before your order can be handled. Please put the completely filled in declaration to the instrument and to the shipping documents in any case. Add also safety sheets and/or specific handling instructions if necessary.
  • Page 108 Mexico – México, D.F Croatia – Zagreb Tel. (3) 94 40 80, Fax (9) 54 80 38 ❑ Endress+Hauser (México), S.A. de C.V. ❑ Endress+Hauser GmbH+Co. Sweden – Sollentuna Tel. (5) 5 55 68 24 07, Fax (5) 5 55 68 74 59 Malaysia –...