Operation With A Non-Hastings Power Supply - Teledyne 200 Series Instruction Manual

Hide thumbs Also See for 200 Series:
Table of Contents

Advertisement

200/202 Series Flowmeters and Controllers Instruction Manual
2.7.2.

Operation with a Non-Hastings Power Supply

See the connection diagrams in sections 2.5.1 and 2.5.2.
WARNING: Care must be taken to avoid any high voltages that may be present when dealing
with power supplies.
To successfully operate a flow controller, a 0-5 VDC command signal proportional to the intended flow (0
Volts = zero flow; 5 Volts = 100% of rated flow) must be applied to pin 14 of the DA-15 (15 Volt), Pin 3 of
the DE-9 (24 Volt) connector. A 0–5 VDC signal proportional to the flow rate through the instrument will be
present on pin 6 of the DA-15 (15 Volt), Pin 2 of the DE-9 (24 Volt) connector. The control mode is selected
via pin 8 of the DA-15 (15 Volt), Pin 6 of the DE-9 (24 Volt) connector. Apply > 10 Volts for full open, < 0
Volts for closed and allow the override to float for flow proportional to the command voltage. Refer to your
power supply manual for the specifics of implementing these parameters.
EAR99 Technology Subject to Restrictions Contained on the Cover Page
Hastings Power Supply Feature Guide
Feature
# of Channels
Front Panel Display
± 15 Volts
Analog Interface
Valve Override
Totalizer
Reads 4-20 mA
4-20 mA Setpoints
Reads/Commands 0-5 VDC
Reads/Commands 0-10 VDC
Ratio Control
Alarms
Communications
THCD-101
THCD-401
1
Yes
Yes
Yes
Yes
No
Yes
No
Yes
Yes
No
Yes
USB emulating RS232, Ethernet
4
Yes
Yes
Yes