Settings; Selection Hart Variable - Siemens SITRANS L Series Operating Instructions Manual

Radar transmitters
Hide thumbs Also See for SITRANS L Series:
Table of Contents

Advertisement

3
Figure 7.11
Quick
commissioning
parameter
Operation
Material type
Measuring
structure
Lower calibration
point
Upper calibration
point
Volume flow unit
Unit
Measurement end
7.7

Settings

7.7.1

Selection HART variable

Secondary value
Sets a process value as secondary variable.
Setting
SELECT OUT
SITRANS LR150
Operating Instructions, 08/2020, A5E50352155
Upper scaling point (0.6 m)
Application example volume flow
Setting/value
Volume flow
Liquid
Liquid
Selection
SELECT OUT
Setup with the integrated HMI
VFLOW
LQD
RWRC
1.6
1
VFLOW UNITS
SCALE FORMAT
UPPER SCALNG
Selection
LEVEL
SPACE
DIST
VOL
7.7 Settings
Description
Venturi flume
Distance to the empty point
or bottom of the measuring
channel. Adjusts the material
level at 4 mA.
Distance to maximum level.
Sets the " Measurement end".
Setting according to the
requirements of the end
user.
Unit which corresponds to
the level (head).
To be taken from the data
sheet of the manufacturer of
the measuring structure.
Description
Level
Space
Distance
Volume