HT and single tank error diagnosis

Tests and measurements in the high voltage circuit

Connect the oscilloscope measurement cables to measuring points HTPOS and HTNEG
or HT on D1.
Adjust the trigger point to the falling edge of the high voltage curve and set the time to
20 - 50 ms/Div.
If the high voltage cannot be displayed, check the tube current at measurement point
MAPOS on D1. Check the inverter current with an external current transformer.
Measure HTPOS and HTNEG on D1. Record the fall times (90 - 10% of HT), high
voltage levels and whether the exposure was interrupted prematurely. Compare with
Fig. 2 "HT and inverter current, inverter and single tank normal" .
Measure the inverter current using an external current transformer and the HT. Compare
with Fig. 1 "HTPEAK and inverter current, inverter and single tank normal" .
In case of an error, compare the oscillogram with Fig. 1 "HTPEAK and inverter current,
inverter and single tank normal" .
NOTE
Siemens AG
Medical Solutions
The inverter current is always measured using a separate current
transformer between the inverter and single tank.
SPR8-220.061.03
Rev. 03
09.03
6
Page 1 of 12
MOBILETT Plus HP
CS PS 24
6 - 1
6