Mechanical systems

Replacing the spring pack

DANGER
Min. 2400
Fig. 10
Equipment required: 2 pieces of wood (50 x 70 x 500 mm), rope, ceiling support and
NOTICE
Ensure compliance with the safety information in the section:
"Safety information and protective measures".
Block the rear wheels and apply the parking brake.
Jack up and block the unit securely 5 cm above the floor. This must be done properly so
that the unit does not tip over.
Siemens AG
Medical Solutions
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
a protective mat.
Difficult operations.
We recommend that two service engineers perform this
replacement.
SPR8-220.061.03
Rev. 03
09.03