Siemens MOBILETT Plus HP Service Instructions Manual page 24

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

2 - 12
Dis-
played
messages
Err 23
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus HP
Mode of
Cause
operation
During exposure The HT inverter has
been interrupted
several times, e.g. by
overcurrent.
SPR8-220.061.03
Rev. 03
Display messages
Possible reasons and advice
When the HT inverter has been inter-
rupted more than 3 times, the expo-
sure will be interrupted.
Faulty inverter/inverter control
Check the inverter current (see
sections "Error diagnosis HT and
single tank", "P12 - Inverter and
single tank test" and "P13 - Inverter
test").
The X-ray tube may have arced
during the exposure. Tube arcs can
occur if the first exposure after a
long interval is an exposure with
high kV values.
The unit has to be started with
several exposures at high mAs
values. The kV values selected
should be increased from 40 kV to
133 kV.
See also section "Error diagnosis
HT and single tank".
Page 12 of 22
09.03
CS PS 24
Siemens AG
Medical Solutions