Siemens MOBILETT Plus HP Service Instructions Manual page 98

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

10 - 14
!
Adjustment screw
Fig. 16
Relieve the spring tension by loosening the adjustment screw (Fig. 16).
Remove all collimator covers and grounding cables.
Disconnect all cables to the collimator.
Remove the collimator according to the information in chapter 9.
Disconnect all electrical connections to the single tank.
Remove the screws and nuts (Fig. 17) bolts placed in the ends of the axis of rotation.
Lift up the fork and remove the single tank. Do not turn the fork around.
The new single tank has to be installed in the same position as the old one.
Replace the locking nuts with new nuts enclosed with the new single tank.
Install the four screws according to Fig. 17.
MOBILETT Plus HP
Mechanical systems
Fig. 17
SPR8-220.061.03
Page 14 of 28
Rev. 03
09.03
CS PS 24
Siemens AG
Medical Solutions