10 - 16
•
Switch on the collimator lamp and check compliance of the light and radiation fields
according to the specifications in the section on "Replacement of collimator".
•
Reattach all covers on the single tank.
•
Adjust the adjustment screw so that the single tank is balanced. If any problems occur
(see section "Adjustment of arm system").
•
Use the service program "P16 - Show the exposure counter" and record the values of the
exposure counter in the service protocol and in the unit documentation.
•
Use the service program "P10 - Fast adaptation" to adjust the filament values to the new
single tank.
•
Reattach the remaining covers.
MOBILETT Plus HP
Mechanical systems
SPR8-220.061.03
Page 16 of 28
Rev. 03
09.03
CS PS 24
Siemens AG
Medical Solutions