P09 - Clear the message history

NOTE
Activate the program with the "lamp" button. The display shows:
kV
P09 CLEAr
To clear the message history, press the "mAs +" button for 5 seconds. The program con-
firms that the history is clear:
kV
0
After the message clears, the program is ended.

P10 - Fast adaptation

This program is used after replacing the single tank or if the memory loses the pro-
grammed values. The user must make a number of exposures.
The generator calculates and stores the optimized filament values for each kV value from
these exposures.
After activating the program, the user must press the exposure button every time the
ready light indicator lights. The program starts at a low kV value and when the actual
value is stable, it increases to the next kV value automatically.
Example:
kV
40
During the adaptation, some messages may appear. Most of these can be ignored since
the generator has incorrect filament values. However, check every message against the
message list and confirm them by pressing the "kV +" button. When the generator is ready
with an adaptation on a kV value, it continues to the next value. When the adaptation is
finished, the program will stop and the display will show:
kV
P10 PASS = Adaptation program completed
MOBILETT Plus HP
When this program is run, the history will be cleared and the file
will be empty. Use program P08 and record all messages before
using program P09.
mAs
mAs
Error = Message history empty
mAs
Ad Pr = Generator in adaptation mode
mAs
SPR8-220.061.03
Rev. 03
Service programs
Page 6 of 8
09.03
CS PS 24
5
5
Siemens AG
Medical Solutions