2 - 2
Dis-
played
messages
Err 4*
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus HP
Mode of
Cause
operation
During
Measured kV value
exposure
deviates by more
than 5 kV from the
selected kV.
SPR8-220.061.03
Rev. 03
Display messages
Possible reasons and advice
The deviation can depend on the
adaptation.
Measure the kV curve at
measuring point "HTPEAK" (see
page 140**). Compare the
measured curve with the reference
curve (see page 141**).
For a large deviation:
Check the filament value (see
section "P07 - Read-out filament
values from battery backed RAM")
and compare it with the values in
the delivery protocol.
If the difference is large a new
adaptation is needed (see section
"P10 - Fast adaptation")
See also section "Error diagnosis
HT and single tank".
Check the connection K9 between
D1 and H1 (single tank) and the
connection between D1 and D13
(K8) (see page 140**)
Page 2 of 22