10 - 20
NOTICE
Set S10 to the OFF position.
Reinstall all covers except single tank covers "A" and "B".
Make sure that the screw mountings between the fork and the arm are tightened
properly.
Remove the rope between the handle and the foot while keeping the single tank on the
table.
Tighten the adjustment screw until the single tank and arm are balanced. Refer to the
section "Adjustment of arm system".
Switch on the collimator lamp and check compliance of the light and radiation fields with
the specifications in the section "Replacement of collimator".
Reattach the remaining covers.
MOBILETT Plus HP
Do not stretch the cable too tightly in the arm.
There must be enough slack so that the cable is not pulled by the
mechanical movements.
SPR8-220.061.03
Rev. 03
09.03
Mechanical systems
Page 20 of 28