In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This can limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-6.
3.3.2.3
RS-232 Port: The digital signal output is a standard RS-232
serial communications port used to connect the analyzer to a computer,
terminal, or other digital device. It requires a standard 9-pin D connector.
The data is status information, in digital form, updated every two
seconds. Status is reported in the following order:
The concentration in percent
The range in use (HI, MED, LO)
The span of the range (0-10 %, etc)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Four input functions using RS-232 have been implemented to date.
They are described in Table 3-1.
Table 3-1: Commands via RS-232 Input
Command
Description
as<enter>
Immediately starts an autospan.
az<enter>
Immediately starts an autozero.
co<enter>
Reports "Raw Cell Output" (current output of the sensor
itself) in µA. For example—
3-10
Figure 3-6: FET Series Resistance
Teledyne Analytical Instruments
Model 3000PA