8.5.26
Shaper 19-27 (2.2.6.3.)
8.5.27
Shaper 28-36 (2.2.6.4.)
8.5.28
Shaper 37-40 (2.2.6.5.)
8.5.29
Rate (2.2.7.)
8.5.30
Response Rate (2.2.7.1.)
Note
Changing Response Rate resets Fill Rate, Empty Rate, Damping Filter and Echo Lock.
Sets the reaction speed of the device to measurement changes.
Related pa‐
Use a setting just faster than the maximum filling or emptying rate (whichever is faster).
SITRANS LR260 with mA/HART
Operating Instructions, 11/2019, A5E32337683-AD
Values
Values
Values
Response Rate
rameters
slow
medium
Options
*
fast
Range: –50 to 50 dB
Default: 0 dB
Range: –50 to 50 dB
Default: 0 dB
Range: –50 to 50 dB
Default: 0 dB
Fill Rate
Empty Rate
0.1 m/min
0.1 m/min
1 m/min
1 m/min
10 m/min
10 m/min
Parameter reference
8.5 Setup (2.)
Damping Filter
Echo Lock
10 s
max. verifica‐
tions
10 s
max verifica‐
tions
0 s
max. verifica‐
tions