Download Print this page

Pulse Xyleject Operator's Manual page 9

Injection device

Advertisement

SECTION 5 – PREPARING THE DEVICE
Needle Selection: Choose the desired gauge of Conical Needle
(18-Gauge or 20-Gauge). Note: Injection instructions and selection
information for needles and injection depth can be found in the
XylPhi-PD Product Insert.
18-Gauge Conical Needle
20-Gauge Conical Needle
Needle Retention Nut Selection: The Needle Retention Nut is attached to the Needle Retention Nut Base and,
together, these components secure the Conical Needle on the injection device. The Xyleject device comes
with a set of Needle Retention Nuts.
Each Needle Retention Nut is optimized for a specific gauge of needle and the desired injection depth.
Choose the appropriate Needle Retention Nut according to the Conical Needle and injection depth
specifications listed in the XylPhi-PD Product Insert.
YELLOW
18-Gauge Needle Retention Nut
4mm injection depth;
BLUE
18-Gauge Needle Retention Nut
6mm injection depth;
SILVER
20-Gauge Needle Retention Nut
3mm injection depth;