SECTION 15 – TROUBLESHOOTING THE INJECTION DEVICE
If the injection device is not dispensing product, check the following:
o Is the device "primed" (i.e. was all air purged from the fluid path during the priming
process)?
o If the device is not primed, refer to the priming instructions in Section 8.
o Does the pressure gauge show sufficient pressure (50 PSI)? If not,
o Check that the Pneumatic Hose is securely attached to the CO
o Check that the CO
o Check that the CO
o Can you depress the trigger easily? Is there a blockage of the orifices of the Conical
Needle?
Note: In examining the Conical Needle, DO NOT bring it close to your eyes!
 If the orifices of the Conical Needle are blocked, then
replace the needle.
 If the orifices are not blocked, observe the Injection
Indication Window to see if the pneumatic piston is
traveling forward. An alternative injection location may
be necessary.
o Is the Medicine Piston O-ring damaged?
o If so, see "Replacing the Medicine Piston O-ring" in Section 14.
If the device is leaking, check the following:
o Is the Holding Compartment securely connected to the injection device?
o If not, finger tighten the Luer Connection.
o If device still leaks, switch to a new Holding Compartment.
o If there is still leakage, switch to a new Luer Connection.
tank is filled.
2
tank valve (24-ounce tank) is open.
2
tank.
2
Injection Indication
Window
28