Download Print this page

Pulse Xyleject Operator's Manual page 11

Injection device

Advertisement

SECTION 7 – FILLING AND INSTALLING THE
HOLDING COMPARTMENT
During the injection process, the injectate is held in the Holding Compartment. (See the XylPhi-PD
Product Insert for information about proper cleanup and disposal in case of any leaks or spills.)
The Holding Compartment must be filled with injectate and air pockets or bubbles must be removed
from the Holding Compartment. Air pockets or bubbles impede the flow of injectate through the
injector.
The following items are used:
Compartment Plunger
Holding Compartment
Suction Tube
Attach the Suction Tube to the Holding Compartment, then screw the Compartment Plunger into the
bottom of the Holding Compartment Plunger, using about 3 turns.
Place the Suction Tube into the Medicine Vial and pull the Compartment Plunger rearward until it
reaches the desired volume level.
11

Advertisement

loading

Related Products for Pulse Xyleject