SECTION 9 – ADJUSTING THE DOSE VOLUME
Move Safety to the Safety ON position. Turn the Dose Setting Knob all the way inward until it stops.
Then turn the Dose Setting Knob
outward (counter-clockwise). The first
full outward turn is for dose volumes up
to 100 microliters ( μl ), using the
reference numbers on the left column.
The second full outward turn is for dose
volumes from 100 to 200 μl , using the
reference numbers on the right column.
For example, if the required dose is 80
μl , turn the Dose Setting Knob in the
direction of the arrow stopping at 080.
15