Maintenance

Kits and Service Options

Agilent pumps will provide many years of trouble-free
service if the maintenance procedures and intervals are
observed. Cleaning and tip seal replacement are recom-
mended when pump base pressure has risen to an unac-
ceptably high level for your application. If your pump
exhibits humming or grinding noises from the bearings, a
major overhaul should be done by Agilent or an autho-
rized rebuild center. Advance exchange pumps are avail-
able to minimize downtime.
The parts needed for tip seal replacement on the IDP-10
are available in the kit described in Table 9. This kit con-
tains seals and O-rings, and can be obtained from your
Agilent dealer.
Table 9 Accessories and Replacement Kits
Part
Description
Number
X3807-67002
Front Cowling
Assembly
X3807-67003
Rear Cowling
Assembly
X3807-68008
Gas Ballast Kit
NOTE
For further service options, please con-
tact your local service center listed in
the back of this manual and ask for your
local representative.
22
IDP-10 Dry Scroll Vacuum Pump
Contents
Replacement front
cowling material to
replace a broken
plastic cowling.
Replacement rear
cowling material to
replace a broken
plastic cowling.
Spare Parts
Service
Agilent offers rebuilding service at our repair centers, or
you can obtain advance exchange of complete IDP-10
pumps. The service plans available and their ordering
numbers are given in Table 10.
Table 10 Service Options
Factory Service Plans
Replacement Tip Seal Set
IDP-10 Exchange Pump
IDP-10 Exchange with Inlet Valve
IDP-10 Full Pump Rebuild
IDP-10 Full Pump (with Inlet
Valve) Rebuild

Cleaning

Exterior
The exterior surfaces of the IDP-10 may be cleaned with a
cloth slightly moistened with water. If necessary alcohol
or a mild detergent may be used instead. Do not pour or
spray any liquids onto the pump.
Interior
Run the pump periodically at atmosphere for a minute or
two to flush it out. For more information, see "Cleaning
the Pump" on page 20.

Tip Seal Replacement

The parts and tools required to replace tip seals include:
Tip Seal Replacement Kit P/N X3807-67000
4 and 8 mm Allen wrenches
Razor blade or side-cutting pliers
Compressed air (optional)
WARNING
If dangerous gases were being pumped,
flush the pump with air or inert gas for
at least 10 minutes prior to disassembly.
Figure 10 on page 23 shows the various components
involved in a tip seal replacement procedure. Refer to
Figure 10 as you follow the procedure.
To remove the worn tip seals:
1. Disconnect the pump from electrical power.
2. Remove (3) M8 socket head bolts (item 1).
Model Number
X3807-67000
X3810-69000
X3810-69010
X3810-69000-RP
X3810-69010-RP