Technical Data Transmitter Sitrans Fus060 - Siemens SITRANS F US Operating Instructions Manual

Ultrasonic flowmeter for use with sensor type sono 3100, sono 3300 and sonokit 1- and 2-tracks
Hide thumbs Also See for SITRANS F US:
Table of Contents

Advertisement

1.6
Technical data - transmitter SITRANS FUS060
Output
Analog output
Digital output 1
Error in measurement:
(at reference conditions)
Digital output 2 (Relay)
Communication via analog output 4...20 mA
Electrical isolation
Ambient temperature (transmitter)
Storage temperature
Enclosure rating
Electromagnetic compatibility
6
Signal range
Upper limit
Signal on alarm
Load
Only PROFIBUS PA version
Active or passive signal, can be configured
with positive or negative logic
For explosion protection
ATEX II 2G Ex dem [ia/ib] T6..T3
Output function, configurable
Pulse output
Analog output
Repeatability
Relay, NC or NO contact
For explosion protection
ATEX II 2G Ex dem
[ia/ib]
IIC T6..T3
Output function, configurable for:
PC/laptop (PDM) or HART communicator
with SITRANS F flowmeter
Load
Cable
Protocol
In potentially explosive atmospheres
Emission
Immunity
SFIDK.PS.029.E5.02 / SW 2.01.07
4...20 mA active current output. (13.2V<open loop voltage<15.8V)
20...22.5 mA, adjustable
3.6 mA, 22 mA or 24 mA
Max 600 ohm for non Ex version
Max 330 ohm Ex version.
Analog output omitted, is replaced by digital PROFIBUS PA
interface
Active: 24 V DC, ≤ 24 mA, R
Passive: open collector, 30 V DC, ≤ 200 mA
Passive: open collector, 30 V DC, ≤ 100 mA
Pulse output
Adjustable pulse significance ≤ 5000 pulses/s
Adjustable pulse width ≥ 0.1 ms
Frequency output
f
selectable up to 10 kHz
END
≤ ± 0.5% of measured value at 0.5 m/s to 10 m/s or
≤ ± 0.25 / v [m/s]% of measured value at flow < 0.5 m/s
As pulse output plus ± 0.1% ± 20 µA
≤ ± 0.25% of measured value at 0.5 m/s to 10 m/s
Switching capacity max. 5 W
Max. 50 V DC, max. DC 200 mA
= 9 Ω
Self-resetting fuse, R
i
Max. 30 V DC, for DC max. 100 mA, for AC max. 50 mA
(cf. EC-Type Examination certificate)
Alarm, flow direction, flow limits, sound velocity limits,
ultrasonic-amplitude limits, no function
Min. load 230 Ohm
2-wire shield ≤ 3 km (≤ 1.86 miles)
Multi-core shield ≤ 1.5 km (≤ 0.93 miles)
HART, version 5.1
Outputs electrically isolated from power supply and
from each other (Output isolation 63 Vac/100 Vdc to ground.)
-20...+50 °C (-4...+122 °F)
Observe temperature class
-25...+80 °C (-13...+176 °F)
IP65
For use in industrial environments
EN 61000-6-3 (Light industry)
EN 61000-6-2 (Industry)
Use shield cables for the outputs
= 300 Ω