Setting The Center Position Of The Cable Module To The Orbital Center Position Of The C-Arm - Siemens POWERMOBIL Manual

Table of Contents

Advertisement

Checks and Adjustments
Fig. 1
Setting the center position of the cable module to the
orbital center position of the C-arm
The lock of the cable module is released, the screw (1/Fig. 2) is in position (2/Fig. 2).
The C-arm can be moved orbitally.
Turn the C-arm orbitally so that the markings (1/Fig. 1) are centered on the threaded
holes (2/Fig. 1). The cable drums are in center position.
Lock the cable module in this position. For this purpose, turn the screw (1/Fig. 2) into po-
sition (1/Fig. 2).
Fig. 3
Loosen the screw (1/Fig. 3), but do not screw it out. Press the screw (1/Fig. 3) in direc-
tion of the arrow (Fig. 3). With this the cable drums are uncoupled from the toothed belt of
the C-arm. The C-arm can be moved slightly orbitally.
Move the C-arm orbitally so that both markings (1/Fig. 4) on the toothed belt are posi-
tioned between the two rollers (2/Fig. 4). The C-arm is now in orbital 0° position.
Siemens AG
Medical Solutions
Fig. 2
Fig. 4
SPR2-240.841.01
Rev. 01
05.03
4