Prerequisite; Safety Information - Siemens POWERMOBIL Manual

Table of Contents

Advertisement

1 - 2

Prerequisite

NOTE
NOTE

Safety Information

WARNING
POWERMOBIL
These instructions apply to the replacement of the cable module
in the POWERMOBIL system.
2 persons are required to carry out any extensive work.
Estimated time required for 2 persons is 3 days.
We recommend that you replace the cable module in a separate
room (workshop).
Please make sure that sufficient working space is available as the
basic unit must be accessible from all sides when replacement
work is carried out.
Avoiding light to severe, possibly fatal injuries and avoiding
material damage.
Product-specific safety information provided in the system man-
ual and the SPR2-240.898.01... service manual as well as the
safety information contained in the ARTD, part 2, must be obser-
ved.
After all work has been completed and all covers have been
attached, perform the protective conductor test according to
ARTD-002.731.17. The protective conductor resistance must not
exceed 0.2 ohms.
Tests and settings that need to be performed under X-radiation
are identified with the radiation symbol
cedures marked with this symbol, radiation protective clothing
must be worn.
SPR2-240.841.01
Rev. 01
Page 2 of 4
05.03
CS PS 24
General
. While performing pro-
Siemens AG
Medical Solutions
1
1