Setting the sensor of the motorized orbital movement

NOTE

Preparations

Make sure again that all cables, plugs, cable clamps and protective ground wires are
connected.
Connect the basic unit with the monitor cart (plug X10).
Disconnect both plugs ST5 and ST6 (1/Fig. 7) from the cable module electronics.
Connect the monitor cart to the power supply.
Switch the system on.
The cable drum is still locked. The screw (1/Fig. 2) is still screwed into position (1/ Fig. 2).

Checking the sensor setting

Loosen the orbital brake by pressing the associated button.
The C-arm can be moved slightly orbitally. At the same time the rollers (2/ Fig.5) move
into the end position inside the bell (1/Fig. 5). In both end positions the LED D3 (2/Fig. 7)
should light up half-brightly.
If the check is not successful, go to the next section "Setting the sensor".
Switch the system off.
Change the position of the stop screw of the cable module back from position (1/Fig. 2) to
position (2/Fig. 2) and tighten it.
Siemens AG
Medical Solutions
The sensor is already adjusted upon delivery of the cable module.
Generally, the setting only needs to be checked according to the
section "Checking the sensor setting".
However, should the sensor not be adjusted properly, it can be
adjusted according to the "Setting the sensor" section.