2 - 2
Determining the electrical I. I. center
•
Attach a washer to the center of the I.I. input. (Remove the grid; the plate with the center
cross may remain attached.)
•
Attach a transparent piece of tape in the center of the monitor.
•
Select the smallest zoom format.
•
Release fluoro.
•
Mark the image of the washer on the adhesive tape.
•
Select full format.
•
Release fluoro.
The washer must appear centered on the marking (smallest format on the adhesive
tape).
•
If necessary, shift the washer on the I.I. input and repeat the entire procedure, until the
"electrical center" is determined.
(tolerance ± 1 mm)
LITHOSTAR Multiline
Centering the X-ray tube / Image intensifier
.
Register
RXL2-120.071.01
Rev. 01
Page 2 of 10
05.97
TD SP 1
2
Siemens AG
Medical Engineering