Disassembly
•
Loosen the POWERPHOS mounting screws (1/Fig. 7). Do not loosen any other screws!
•
Loosen all POWERPHOS plug connections.
•
Loosen the two inverter connections (terminal screws).
•
Loosen the POWERPHOS protective ground wire.
•
Loosen the two clamps on the front of the POWERPHOS.
NOTE
•
Lift the POWERPHOS out of the C-arm and set it down.
•
Eject all labeled contact pins from the Z66 X1 plug (Fig. 8).
•
Remove any existing cable ties.
Siemens AG
Medical Solutions
To facilitate lifting the POWERPHOS out of the C-arm, thread
2 cable ties (min. 60 x 6 mm) into the existing holes.
SPR2-240.841.01
Rev. 01
Page 5 of 12
05.03
CS PS 24
2 - 5
POWERMOBIL