3 - 4
Fig. 9
Fig. 11
Fig. 9 shows the horizontal carriage in its end position (moved out fully).
Adapt the cable lengths in the horizontal carriage as illustrated in Fig. 9. For this pur-
pose, push any excess lengths down into the vertical tube. Some cables have a red mark
running round them. They lie in the cable clamp (1/Fig. 10).
Reconnect the plug connections (1/Fig. 9).
Reattach cables and plugs using cable ties (2/Fig. 9).
Refit the cable guide (1/Fig. 11) and cable clamps (1/Fig. 10) and (3/Fig. 9).
Move the horizontal carriage from end position to end position several times. At the same
time, check that the cables run neatly and the loops forming in the end positions are nei-
ther too short nor too long. If necessary, adapt cable lengths once again.