When running cables along the frame for the PC boards D1 and D30 (Fig. 22), make
sure that they are not crushed when inserting and tightening the frame.
Fix the run cables using cable ties.
Refit the plate (3/Fig. 21).
Fig. 23

C-arm cover

Secure the front C-arm cover to the image intensifier side of the C-arm (Fig. 23).
When inserting cables make sure that they are not crushed or torn off.

Covers of the horizontal carriage

Refit the cover plate (2/Fig. 24).
Reattach the cable in the cable clamps. Make sure that the shielding makes good ground
contact.
Screw the protective ground wire to the cover of the horizontal carriage.
Reconnect the plugs of the brake button and of the Emergency Stop switch.
Lift the cover above the horizontal carriage and fasten it with screws (1/Fig. 24).
POWERMOBIL
Fig. 24
SPR2-240.841.01
Rev. 01
05.03