Language Reference
STATus Subsystem
Factory Preset
and *RST:
Range:

Questionable Integrity Signal Positive Transition

:STATus:QUEStionable:INTegrity:SIGNal:PTRansition <number>
:STATus:QUEStionable:INTegrity:SIGNal:PTRansition?
This command determines what bits in the Questionable Integrity
Signal Condition register will set the corresponding bit in the
Questionable Integrity Signal Event register when that bit has a
negative transition (1 to 0). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Factory Preset
and *RST:
Range:

Questionable Negative Transition

:STATus:QUEStionable:NTRansition <number>
:STATus:QUEStionable:NTRansition?
This command determines what bits in the Questionable Condition
register will set the corresponding bit in the Questionable Event
register when that bit has a negative transition (1 to 0). The variable
<number> is the sum of the decimal values of the bits that you want to
enable.
Factory Preset
and *RST:
Range:

Questionable Power Condition

:STATus:QUEStionable:POWer:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Power Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
276
0, factory default. (The user setting is persistant.)
0 to 32767
32767 (all 1's), factory default. (The user setting is
persistant.)
0 to 32767
0, factory default. (The user setting is persistant.)
0 to 32767
Chapter 5