Display Subsystem; Display Annotation Title Data; Display Annotation Title On/Off; Turn The Whole Display On/Off - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

DISPlay Subsystem

The DISPlay controls the selection and presentation of textual,
graphical, and TRACe information. Within a DISPlay, information may
be separated into individual WINDows.

Display Annotation Title Data

:DISPlay:ANNotation:TITLe:DATA <string>
:DISPlay:ANNotation:TITLe:DATA?
Enters the text that will be displayed in the user title area of the
display.
Front Panel
Access:
Display, Title, Edit Title

Display Annotation Title On/Off

:DISPlay:ANNotation:TITLe[:STATe] OFF|ON|0|1
:DISPlay:ANNotation:TITLe[:STATe]?
Turns the display of the title annotation on or off.
Front Panel
Access:
Display, Title, Title On Off

Turn the Whole Display On/Off

:DISPlay:ENABle OFF|ON|0|1
Controls the updating of the display. If enable is set to off, the display
will appear to "freeze" in its current state. Measurements will run
faster if the instrument doesn't have to keep updating the display.
Factory Preset
and *RST:
On

Select Display Format

:DISPlay:FORMat:TILE
Selects the viewing format that displays multiple windows of the
current measurement data simultaneously. Use DISP:FORM:ZOOM to
return the display to a single window.
Front Panel
Access:
Zoom
Chapter 5
Language Reference
DISPlay Subsystem
173